4. Sterilization, Replacement Parts, and Storage
•
To prevent the spread of serious, life-threatening infections, the handpiece grips and its hanger, tips
and tip stand must be cleaned and sterilized between patients.
•
All the handpiece grips and its hanger, tips and tip stand are delivered in
non-sterile condition.
Clean and sterilize them prior to initial use.
•
Always turn off the key and the circuit protector before cleaning. This will avoid the risk of burns
and electric shocks as well as accidents that could results from accidentally pressing a switch.
<All information about cleaning and sterilization (Autoclaving) >
Cleaning and Sterilization
Detail
Tips
Immerse the tip fiber in tap water and emit the laser.
Use the enzymatic detergent (CIDEZYME Johnson & Johnson company: for
example) provided to clean off blood and other contaminants.
Immerse the tip in an available chemical disinfectant
・
Chlorhexidine gluconate
・
Ethanol for disinfection (Ethanol 70 vol% to 80 vol%)
Autoclaving (+135°C (+275°F) 10 to 15 min)
Refer to
Handpiece
grips
Wipe entire handpiece grips outer surface with a soft cloth and ethanol for
disinfection (Ethanol 70 vol% to 80 vol%).
Autoclaving (+135°C (+275°F) 10 to 15 min)
Refer to
Hanger
Wipe the hanger with a soft cloth and ethanol for disinfection (Ethanol 70 vol% to
80 vol%).
Autoclaving (+135°C (+275°F) 10 to 15 min)
Refer to
Tip stand
Wipe the tip stand with a soft cloth and ethanol for disinfection (Ethanol 70 vol%
to 80 vol%).
Autoclaving (+135°C (+275°F) 10 to 15 min)
Refer to
Cleaning only
Detail
Main unit
Wipe the outside of the main unit with ethanol for disinfection (Ethanol 70 vol% to
80 vol%) or neutral detergent
Refer to
* We recommend autoclaving time longer than 10 minutes.