Instructions for Use 2018-04-23
42
(1) Handpiece grips, Hanger, Tips and Tip stand Cleaning and Sterilization
1) Cleaning (Always perform this procedure prior to sterilization)
The cleaning process is intended to remove blood, protein and other potential contaminants from tips,
handpiece grips and hanger. This will not sterilize them.
Contamination control should be performed by
trained personnel, while wearing protective gear (including masks gloves and shields).
<Tip>
(1) After use the tip, immerse the fiber part of the tip in tap water and emit the laser for 3 to 5 seconds.
Use the enzymatic detergent (CIDEZYME Johnson & Johnson company: for example) provided
to clean off blood and other contaminants.
(2) Immerse the tip in an available chemical disinfectant for recommended time by the disinfectant’s
manufacturer’s directions.
Use one of the disinfectants listed below at the concentration specified for medical instruments for
cleaning solution.
· Chlorhexidine Gluconate (Hibiten, for example)
· Ethanol for disinfection (Ethanol 70 vol% to 80 vol%)
· Do not use ultra sonic cleaning methods; this could break the tip or cause its connector to
fall off.
· Contact tips will wear out and must be replaced periodically. Replace damaged tips and
those that cannot be adequately cleaned.
(3) Wash tips thoroughly with tap water after using the cleaning solution.
(4) Wipe the contact tip with cotton. Take care not to pull out the h
ollow waveguide
.
<Handpiece grips>
Wipe entire handpiece grips outer surface with a soft cloth
dampened with ethanol for disinfection (Ethanol 70 vol% to 80
vol%).
<Hanger>
(1) Take the hanger off its arm.
(2) Wipe entire hanger outer surface with a soft cloth dampened
with ethanol for disinfection (Ethanol 70 vol% to 80vol%).
<Tip stand>
Wipe entire tip stand outer surface with a soft cloth dampened
with ethanol for disinfection (Ethanol 70 vol% to 80 vol%).
Handpiece grips
Hanger