Stroke volume range
(typical values)
Adult
using one hand
800ml
Adult
using two hands
1350ml
Child
350ml
Infant
100ml
PERFORMANCE CHARACTERISTICS
The performance characteristics for Manual Resuscitators will vary from
user to user depending on a variety of factors: ambient temperature,
patient lung compliance, ventilator frequency, size of operator's hands.
The following data has been derived from Manual Resuscitators in accor-
dance with ISO standards.
Temperature
-18°C/0°F 22°C/72°F 50°C/122°F
Cycle rate
Adult
38
45
45
Child A
92
105
105
Child B
72
78
76
Infant A
95
95
95
Infant B
60
60
60
These results were obtained under the following conditions:
Adult VT-600mlcompliance -0,02 1/cmH
2
O, resistance
-20
cmH
2
O(1/S)
Child A
VT-70ml compliance -0,01 1/cmH
2
O, resistance
-20
cmH
2
O(1/S)
Child B
VT-300ml compliance -0,02 1/cmH
2
O, resistance
-20
cmH
2
O(1/S)
Infant A
VT-20ml compliance -0,001 1/cmH
2
O,
resistance -400 cmH
2
O(1/S)
Infant B
VT-70ml compliance -0,01 1/cmH
2
O,
resistance -400 cmH
2
O(1/S)
PERFORMANCE CHARACTERISTICS
ADULT
Values in parentheses are without an oxygen reservoir
O
2
flow in Tidal vol. (ml) x Ventilation rate
1/min.
600x12 600x20 750x20
750x20
1000x12
1000x20
05
82(33)
58(33)
65(33)
99(30)
99(30)
85(33)
10
99(37)
80(37)
99(37)
99(37)
99(37)
99(37)
15
99(45)
99(45)
99(45)
99(45)
99(45)
85(45)
Conditions:
compliance -0,02 1/cmH
2
O,
Resistance -20 cmH
2
O/(1/S) I:E ratio 1:2
24
CHILD
Values in parentheses are without an oxygen reservoir
O
2
flow in Tidal vol. (ml) x Ventilation rate
1/min.
70x20
70x30
200x20
200x30
300x20
300x30
02
99(65)
99(65)
58(37)
50(37)
45(34)
40(32)
05
99(86)
99(83)
99(44)
96(43)
84(37)
68(36)
10
99(93)
99(90)
99(46)
99(46)
99(45)
99(43)
Conditions:
Vt >150 <600ml
compliance -0,02 1/cmH
2
O,
Resistance -20 cmH
2
O/(1/S) I:E ratio: 1:2
INFANT
Values in parentheses are without an oxygen reservoir.
O
2
flow in Tidal vol. (ml) x Ventilation rate
1/min.
20x30
20x60
40x30
40x60
70x30
70x60
02
99(72)
99(70)
99(70)
99(58)
99(55)
85(50)
05
99(77)
99(75)
99(74)
99(72)
99(65)
85(62)
10
99(94)
99(90)
99(90)
99(80)
99(79)
99(71)
Conditions:
V
T
=20ml
Compliance -0,001 1/cmH
2
O,
Resistance -400 cmH
2
O/(1/S) I:E ratio: 1:1
V
T
=40ml & 70ml
Compliance -0,01 1/cmH
2
O,
Resistance -20 cmH
2
O/(1/S) I:E 1:2
Inspiratory flow resistance
Expiratory flow resistance
ENGLISH
ENGLISH
25
R
esis
tanc
e (
cmH
2
O
/L/
S
)
Flow (LPM)
Flow (LPM)
R
esis
tanc
e (
cmH
2
O/
L
/S
)
08 Mod. 09-2002 BILINGUE M.palloni 8-03-2006 16:44 Pagina 13