background image

status 03/06 - PVE41XGB.doc

 

page 6 of 12 

subject to alterations; errors including printer’s errors

 

Notes on programming 

 

The following programming examples describe the two different tuning methods for setting the 
device. A brief explanation will first be given on documenting the display procedure after pressing the 
[P] key. 
 
If the keyboard lock is not set (see programming lock), pressing the [P] key will always switch to 
programming mode with the program number 0. For approx. 3 seconds, a 0 will appear in the display, 
preceded by a 

P

 – see programming examples. After 3 seconds, the calibration mode 0 will flash 

alternately with the program number 0 for a further 4 seconds. By pressing the 

W

 or 

V

 keys, the 

value stored for the calibration mode is displayed for approx. 3 seconds, during which it can be 
changed with the 

W

 or 

V

 keys. After the 3 seconds have elapsed, the program number 0 flashes 

alternately with the currently set calibration mode for further 4 seconds. The changed value can be 
saved by simultaneously pressing the [P] and 

W

 keys simultaneously. The device acknowledges this 

by displaying 4 horizontal bars. Changing to program number 1 is done by pressing the [P] and 

V

 

keys. 
 
All further settings can be made by following the above procedure. If you are in programming mode 
and do not press any key within 7 seconds, the device automatically reverts from the programming 
mode to the operating mode. You can change back again to programming mode at any time by 
pressing the [P] key. 
 

Programming example 

 

1. Sensor calibration / Tuning the measuring section to a real measuring value 

 
When using the sensor calibration, a voltage or current signal must be applied to the display unit for 
the scaling. Furthermore, it must be ensured that the device is supplied with the correct auxiliary 
voltage. During 

sensor calibration

, the device is tuned to a 

real measuring value

 
The base for this programming example is the basic value after resetting to the factory settings. 
 
For the sensor calibration, a zero must always be stored under program number 0!  
 
Measuring input: 

0/4...20 mA 

Measuring signal:  4...20 mA 
Display: 

 

4 mA = 0.0 

   20 

mA 

300.0 

Display time:  

2.0 seconds 

Set point S1:  

60.0 and closed circuit 

 

 

 

falling at 60.0 and rising at 58.0 

   (corresponding 

hysteresis 

2.0) 

Set point S2:  

150.0 and open circuit 

 

 

 

Rising at 150.0 and falling at 80.0 

   (corresponding 

hysteresis 

70.0) 

Analogue output:  Measuring signal 4 mA corresponds to a display of 0.0 and 0 V at the analogue  
 output. 
 

 

 

Measuring signal 20 mA corresponds to a display of 300.0 and  
10 V at the analog output. 

 

 

Содержание PVE 4.001.1422B

Страница 1: ...segment display Setpoint indication 1 Setpoint indication 2 Insertable dimensionstrip Terminal connection 10V 0 4 20mA 0V sensor supply option analog output option 230V AC 115V AC 15 13 12 11 10 9 8 7...

Страница 2: ...22B Power supply 115 VAC terminal connection 11 L 10 N PVE 4 001 6722B Power supply 24 VDC terminal connection 11 L 10 L galvanic insulated TYPE ORDER NUMBER housing size 48x24 PVE 4 001 7782B Power s...

Страница 3: ...ayed value by pressing the W or V key 7 With program numbers 1 and 2 the applied voltage sensor calibration under program number 0 active can be saved by simultaneously pressing the P and W keys This...

Страница 4: ...de MIN MAX data Additional functions in normal mode for memory inquiry of the MIN MAX data The MIN MAX memory is a volatile memory in which after switching on the device or since the last erasure the...

Страница 5: ...nt no decimal point 4 Input of display time Display time measuring time Integrated measuring process 0 1 10 0 seconds 1 0 Analogue output optional 5 Input of final value for analogue output e g 300 0...

Страница 6: ...and V keys All further settings can be made by following the above procedure If you are in programming mode and do not press any key within 7 seconds the device automatically reverts from the programm...

Страница 7: ...sal bars Offset To program number 2 with P and V Connect measuring value 4mA To memorized value with W or V Memorize value with P and W Take over by display of transversal bars Decimal point To progra...

Страница 8: ...nciple S2 To program number 68 with P and V To memorized value with W or V Programming terminated 2 Factory calibration standard signals When using the factory calibration there is no need to apply a...

Страница 9: ...10 V at the analogue output Start of programming Switch on supply voltage Segment test Operating mode Calibration mode Enter program mode To memorized value with W or V Select parameter 1 for input 0...

Страница 10: ...230VAC 115VAC 24VDC Transmitter Input 0 10V Sensor supply minus Sensor supply plus Notes factory settings and error elimination After the supply voltage is switched on the device performs a reset inc...

Страница 11: ...V 0 20 mA 4 20 mA Range 72x36 0 10 V 0 50 V 0 20 mA 4 20 mA Range 48x24 0 10 V 0 50 V 0 20 mA 4 20 mA The maximum permitted value on the input clips is is 120 of the nominal value All ranges can be s...

Страница 12: ...een only for 96x48 4 digit indication 9999 digit Overflow indication of 4 transversal bars Display time adjustable from 0 1 10 0 seconds Ambient Working temperature 0 60 C conditions Storing temperatu...

Отзывы: