Page 12

I.

Speed of Response:

1. Check that the probe to test plate spacing is still 1/8” (3,2mm).

2. Set the gain pot, front panel recessed, for about midrange.

3. Connect the oscilloscope to the rear panel output connector.

4. With the  279  in  standby,  STBY/OPER switch  out,  check  the  range  of  the  ZERO  2 pot on the front  panel 

meter.  It should swing plus and minus with a total of about 10V.  Zero the front panel meter with ZERO 2.

5. Push the operate switch and then zero the front panel meter with the ZERO 1 pot on the front panel.

6. Apply a repe1kV pulse to the test plate.

7. Adjust the gain pot, the integrator gain pot (R54) and the oscilloscope controls until a critically damped step 

response  between 0% and 100% is observed.  Rise and fall times  from 10% to 90% should be 2.5ms or 
less.

8. Repeat for +2kv.  The 10% to 90% response time should be 3.5ms or less.

9. Repeat for +3kv.  The 10% to 90% response time should be 4ms or less.

10. Check the response to negative steps in a similar manner.  Switch to standby.

J.

Noise Check:

1. Set the probe to test plate spacing to 0.005” (0,13mm).

2. Reduce the gain to just up from counter clockwise.  Switch to operate.

3. Pulse the test plate with +100 volt pulses.

4. Adjust the gain and oscilloscope controls for a critically damped transient response.

5. Replace the oscilloscope with an rms voltmeter.  Noise is typically 0.3mV rms.

K.

Output Calibration:

1. Connect the rear panel output connector to a DVM

2. Adjust ZERO 1 for zero volts on the DVM and the front panel meter.

3. Slowly apply voltage to the test plate until it is pre3.000kV.

4. The DVM at the output 3.000V 

0.1%.  Adjust R5 if necessary to meet this condition.

5. The front panel meter should in3000 

0.1% LSD.

6. Repeat steps 3 through 5 using –3000v.

7. Check calibration at 

2kV.

8. Remove the test equipment and replace the cover.  This completes the adjustment procedure.

Содержание ISOPROBE 279

Страница 1: ...INSTRUCTION MANUAL ISOPROBE ELECTROSTATIC VOLTMETER MODEL 279 Equipment Included Line Cord Manual Probe Cover Plug 279 100 P N 0340112 110309...

Страница 2: ...the non conforming goods and or defective work and in accordance with the foregoing Monroe Electronics shall not be liable for any other damages either direct or consequential RETURN POLICIES AND PROC...

Страница 3: ...I CONTROLS AND INDICATORS PAGE 5 IV OPERATION PAGE 7 V THEORY PAGE 8 VI ADJUSTMENT PAGE 10 VII TROUBLESHOOTING PAGE 14 APPENDIXES MODEL 1034 PROBE MOUNTING APPENDIX I Page 16 HV BREAKDOWN CONSIDERATIO...

Страница 4: ...lectret research Static electrification electric field studies Process monitoring and control B Features Low profile half rack size for multi channel applications Full complement of interchangeable pr...

Страница 5: ...in a normal laboratory atmosphere Separation for accuracy and response speed tests is 1 8 3mm and for noise and drift tests 0 005 0 13mm Performance generally improves in controlled environments and m...

Страница 6: ...he surface being measured Further probe mounting information is given in Appendix I B Probe to surface spacing CAUTION When operating at probe to surface spacing of less than 0 125 inches 3mm BEWARE O...

Страница 7: ...re should be taken to maintain the purged air as free from chemical vapor contamination as possible Probes may also be purged using clean dry oil free air from a local source or through the use of dry...

Страница 8: ...tential to obtain probe to surface spacing insensitivity This adjustment procedure is described in Section 4A ZERO 1 may also be used for large zero offset when spacing insensitivity is not needed D G...

Страница 9: ...C jack is a low voltage representation of the input The output voltage is equal to the measured voltage measured divided by 1000 or 0 to 3 volts for full scale 17 volts full scale typical for 200 opti...

Страница 10: ...witch into OPERATE mode 3 Adjust ZERO 1 for an indication of 0000 on meter 4 Proceed with measurements B Restoration of factory settings The following steps are necessary only if the factory settings...

Страница 11: ...utput of the phase sensitive detector feeds a DC integrating amplifier Its output polarity is inverted to that of the unknown The output of this amplifier is used to drive a high voltage amplifier whi...

Страница 12: ...electrode C Zero ZERO 2 The instrument s zero control is a voltage source connected in the metering circuitry It is used to overcome offsets produced by voltage sources external to the probe D Gain Th...

Страница 13: ...used at JP8 C Initial Setup and Adjustments 1 Insure that available nominal power line voltage and frequency match those indicated of the back panel of the unit Do not apply power at this time 2 Remo...

Страница 14: ...e power is on high voltage will be present on the high voltage rectifier and at the rear of the FET boards 6 Switch on the rear panel power switch 7 Adjust R55 for 40mV on the DVM 8 Move the DVM input...

Страница 15: ...uld be 3 5ms or less 9 Repeat for 3kv The 10 to 90 response time should be 4ms or less 10 Check the response to negative steps in a similar manner Switch to standby J Noise Check 1 Set the probe to te...

Страница 16: ...GAIN R55 BIAS R56 BIAS R65 DEMODulator PHASE R107 GAIN Recessed through front panel R108 ZERO 1 10 turn pot on front panel with knob R111 ZERO 2 Recessed through front panel TP1 Located on FET boards...

Страница 17: ...hecklist 1 INDICATION Power switch at rear of instrument is ON no display illumination a Check display button recessed behind front panel Should be latched in to turn display on See Paragraph 3B b Che...

Страница 18: ...p p sine wave of about 1kHz present The amplitude should be adjustable with R 53 Trouble here usually indicates a defective probe 2 Move the oscilloscope probe to TP 3 Turn the front panel GAIN pot R...

Страница 19: ...nd that the probe assumes the potential of the surface under measurement within these limits The probe therefore must be insulated from other parts of the system Choice of insulating materials althoug...

Страница 20: ...Page 17 Figure A I 1...

Страница 21: ...al voltage to which the power entry module is set is indicated by the indicator pin visible in one of the holes adjacent the voltage markings To change selected voltage open cover using a small blade...

Страница 22: ...ol loosen Phillips screw 2 turns remove fuse block by sliding up then away from Phillips screw and lifting up from pedestal change fuses note that two European fuses are required although a dummy fuse...

Страница 23: ...nder test and or the sensitive circuitry of the probe Under normal operating conditions this is usually not a problem as the probe housing is driven to a voltage essentially identical to the unknown a...

Страница 24: ...directly to the sensitive electrode The influence of these voltage sources is usually not the same at all probe to surface separations Their influence will vary in proportion to the probe to surface...

Отзывы: