Page 12
I.
Speed of Response:
1. Check that the probe to test plate spacing is still 1/8” (3,2mm).
2. Set the gain pot, front panel recessed, for about midrange.
3. Connect the oscilloscope to the rear panel output connector.
4. With the 279 in standby, STBY/OPER switch out, check the range of the ZERO 2 pot on the front panel
meter. It should swing plus and minus with a total of about 10V. Zero the front panel meter with ZERO 2.
5. Push the operate switch and then zero the front panel meter with the ZERO 1 pot on the front panel.
6. Apply a repe1kV pulse to the test plate.
7. Adjust the gain pot, the integrator gain pot (R54) and the oscilloscope controls until a critically damped step
response between 0% and 100% is observed. Rise and fall times from 10% to 90% should be 2.5ms or
less.
8. Repeat for +2kv. The 10% to 90% response time should be 3.5ms or less.
9. Repeat for +3kv. The 10% to 90% response time should be 4ms or less.
10. Check the response to negative steps in a similar manner. Switch to standby.
J.
Noise Check:
1. Set the probe to test plate spacing to 0.005” (0,13mm).
2. Reduce the gain to just up from counter clockwise. Switch to operate.
3. Pulse the test plate with +100 volt pulses.
4. Adjust the gain and oscilloscope controls for a critically damped transient response.
5. Replace the oscilloscope with an rms voltmeter. Noise is typically 0.3mV rms.
K.
Output Calibration:
1. Connect the rear panel output connector to a DVM
2. Adjust ZERO 1 for zero volts on the DVM and the front panel meter.
3. Slowly apply voltage to the test plate until it is pre3.000kV.
4. The DVM at the output 3.000V
0.1%. Adjust R5 if necessary to meet this condition.
5. The front panel meter should in3000
0.1% LSD.
6. Repeat steps 3 through 5 using –3000v.
7. Check calibration at
2kV.
8. Remove the test equipment and replace the cover. This completes the adjustment procedure.