Monroe Electronics 287a Скачать руководство пользователя страница 6

IPA287Supplement Page 

5

 

Monroe Electronics, Inc

 

 
 
 
 
 

Auto Sequence Group Test 

(press [AUTO] button less than 2 seconds)

 

The 287 unit will perform up to 10 groups of automatic test sequences and store the results in memory.  
Groups are named 1 thru 0 where 0 is the 10

th

.  Each group has 3 +decay, 3-decay, and an optional balance 

test. The sequence can be set in the Set-Up Menu.

 

Decay test

 

: Zeroing to <5V, Charge to over +/-1050V, then Decay.

The decay timer starts at +/-1000V and stops at +/-100V. 

Balance Test:

 

 Zeroing to <5V, then Balance for the time duration. Start recording 

 

Voltage peaks after 5 seconds.

The delay time between the tests can be set in the Set-Up Menu.

---V   A30%M50%

28C 9.0V 34%RH  

Main entrance display

AUTO

 

28C 9.0V 34%RH  

 

Null

 

28C 9.0V 34%RH  

Press [AUTO] and hold

less than 2 seconds

 

longer than 2 seconds

+   1V   Zeroing

 

28C 9.0V 34% 11+  

 

+ 950V   +Charge

 

28C 9.0V 34% 11+  

 

Timer starts at 1000V
and stops at 100V.

Group #1 (1-0)
Test #1 (1-3)
+Decay

+   1V   Zeroing
28C 9.0V 34% 1Ba  

Group #1 (1-0)

 

Balance test

 

+   1V    5.2s
+0002mx –0004mn  

 Voltage update   Balance duration timer
  Max and min voltages start recording
  after 5 seconds.

Press [Enter] will abort any on going test and back to 

Main

.

After the test complete, review data begins.

 

Press [^] or [v] will scroll through the data.

 

Press [Enter] or [ON/OFF] back to 

Main

 

Auto Sequence Test

 

go on for the other 5 decay tests...

 

Review test results

+1030V   +Decay

 

28C 9.0V 34% 11+  

 

+ 800V     3.6s

 

28C 9.0V 34% 11+  

 

+  20V   8.6s

 

Delay   6 sec  

 

Содержание 287a

Страница 1: ...Supplementary Manual for Model 287a Ionizer Performance Analyzer 287a 1 100A V1 12 P N 0345006 101107...

Страница 2: ...or v Press or v Press or v Press or v Press or v Press or v Press or v Press or v Press Enter ClrData N Y N A30 M50 Press Enter Memory used Auto sequence tests 30 Manual tests 50 If no data in memory...

Страница 3: ...v 0005v 1TH 25C 30 RH Press or v M2D 7 3s Manual test Test 2 1 0 Decay Decay test time in seconds M1D 6 8s Manual test Test 1 1 0 Decay Decay test time in seconds M3B 0012v 0003v Manual test Test 3 1...

Страница 4: ...art recording after 5 seconds Press Enter will abort any on going test After the test complete review data begins Press or v will scroll through the data Press Enter or ON OFF back to Main Manual Plus...

Страница 5: ...ge update Balance duration timer min and max voltages start recording after 5 seconds Press Enter will abort any on going test After the test complete review data begins Press or v will scroll through...

Страница 6: ...the tests can be set in the Set Up Menu V A30 M50 28C 9 0V 34 RH Main entrance display AUTO 28C 9 0V 34 RH Null 28C 9 0V 34 RH Press AUTO and hold less than 2 seconds longer than 2 seconds 1V Zeroing...

Страница 7: ...charge to 1050V Make sure nothing is contacting the charge plate and the standoffs underneath are clean This may also be caused by too much load on the plate Press enter to return to the main screen...

Отзывы: