background image

 

 

Maintenance 

 

Replacing the battery 

The battery should be replaced annually, whenever you plan an extended period 

of unattended monitoring, or whenever “BAT” appears in the display for more than an instant. Low BAT is 
set for 6.9 

±

 0.6 volts.  Remove the battery before long-term storage. See Specifications for replacement 

type.  If unit fails to operate after replacing the battery, remove battery, press and release the ON/ZERO 
button then re-connect the battery.  

Cleaning 

To obtain accurate, drift-free readings the sensor plate and especially the area

 around the 

aperture must be kept absolutely clean at all times. Be sure to always keep the instrument in its protective 
case when not actually taking measurements.  

Never touch the aperture with anything-not even 

cotton swabs. 

 To remove dust or other particulate matter, use low-pressure instrument grade air.  To 

remove severe contamination spray or flush with the smallest practical amounts of clean technical grade 
isopropyl alcohol.  The stand the instrument on end and allow it to dry for several hours. 

DO NOT 

clean 

with any solvent on a regular basis. 

Zero Drift  

Apply anti-stat solution to plastic surrounding the sensor plate using lint free applicator.  

DO 

NOT

 allow entry into aperture. 

 

 

Calibration 

 

Monroe Electronics instruments are factory-calibrated prior to shipment. Recalibration should be 
performed annually, or more frequently if specified by contract or company policy. Your instrument should 
also be recalibrated any time it has been repaired or tampered with. 
We are happy to recalibrate your instrument for you at a reasonable cost, or provide information and 
procedures on calibration upon request. We also offer calibration services with traceability to NIST or MIL-
STD-45662A. 
 

Safety Considerations 

 

The Model 282A-1 has undergone testing to qualify for the following markings: 
 

CE

 — Radiated and ESD immunity to severity levels for industrial devices according to EN 50082-1 and 

radiated emissions with limits for Group 1, Class A product of EN 55011 to the generic standard EN 5008 
1-2 (1992). 
 

Two-year warranty 

 

Monroe Electronics, Inc. warrants this instrument to be free from defects in material and workmanship for 
a period of two years after shipment from the factory. This warranty is applicable to the original purchaser 
only. Liability is limited to service, adjustment or replacement of defective parts (other than fuses or 
batteries) on any instrument returned to the factory, transportation charges prepaid. It does not apply to 
any instrument subjected to abuse, abnormal operating conditions, or unauthorized repair or modification. 
No other warranties, explicit or implied, apply. Under no circumstances shall Monroe Electronics be 
responsible for damages either direct or consequential. For additional information concerning this 
product, contact Monroe Electronics, Inc.

 

 
 
 
 
 
 
 
 

Copyright

 1999 Monroe Electronics, Inc. 

Printed in USA 

Stat-Arc is a trademark of Monroe Electronics, Inc. 

Specifications subject to change without notice. 

 