Main ULPA Filter (VLFT units)
Open the control panel by removing the 2 screws inside the lower cabinet.
Remove the two filter clamp bolts.
Lift and hold open the head assembly using the prop at the right hand side.
The ULPA filter may now be withdrawn from the front of the cabinet and a
replacement fitted.
The replacement filter should be DOP tested prior to use and the airflow and
alarm re-calibrated if necessary.
Main ULPA Filter (HLFT units)
Remove the screws securing the back panel and lift it clear.
Remove the screws securing the filter clamp frame.
The HEPA filter may now be withdrawn from the rear of the cabinet and
a replacement fitted.
The replacement filter should be DOP tested prior to use and the airflow and
alarm re-calibrated if necessary.
Model
Part #
HLFT 900
K-HF0139
HLFT 1000
K-HF0140
HLFT 1200
K-HF0141
HLFT 1500
K-HF0142
HLFT 1800
K-HF0143
Model
Part #
VLFT1000
K-HF0135
VLFT 1200
K-HF0136
VLFT 1500
K-HF0137
VLFT 1800
K-HF0138