INTRODUCTION 

Monmouth Scientific Clean Air Module (CAM) is a self-contained filtration unit 
designed to fit into a ceiling to provide up to 1000m³/hr of HEPA filtered clean air.  

 
 

INSTALLATION AND SETUP 

 
The module requires a hole cut out of 970mm x 665mm and will sit flush in an 
aperture thickness of 50mm. 
 
It should also be suitably fixed using the holes on the top angle brackets. 
 

ELECTRICAL CONNECTION 

 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
The module requires a 230vac electrical supply and can be supplied via a domestic 
13A socket outlet or fused spur.  
 
A power-through auxiliary IEC socket is provided for the connection of lighting. 
Please ensure the current load of 3A is not exceeded. 

(Note: The module provides no control over power from this socket. If switched lighting is required for 
example then an additional provision will need to be externally provided.) 

 
Individual fuses are provided for circuit protection of the Fan and Aux Lighting power 
outlet. 
 
 
 
 
 
 
 
 
 
 
 
 

Power inlet socket 

Power switch 

Aux Light Power 

Outlet 

Содержание CAM1000

Страница 1: ...Operating Maintenance Manual www monmouthscientific co uk Clean Air Module CAM1000 ...

Страница 2: ...ly exploited in any form or by any means redistributed or reproduced in entirety or in part in any form for individual or third party use photocopied or stored on any form of electronic retrieval system or linked to any other website without the prior express written permission of Monmouth Scientific Limited For permission requests please write to Monmouth Scientific Limited Units 5 5 Kilnside Eas...

Страница 3: ...ICAL CONNECTION The module requires a 230vac electrical supply and can be supplied via a domestic 13A socket outlet or fused spur A power through auxiliary IEC socket is provided for the connection of lighting Please ensure the current load of 3A is not exceeded Note The module provides no control over power from this socket If switched lighting is required for example then an additional provision...

Страница 4: ... required to achieve N air changes per hour can be calculated by the following formula Air volume required m hour Room Volume m X N The air volume through the module can be calculated by the following formula CAM1000 Air Volume m hour 0 37 filter area in m X measured HEPA velocity m sec x 3600 secs hr Based on the filter size used in this module 610mm x 610mm the following table gives an indicatio...

Страница 5: ...rame 3 Remove the filter frame cover filter if fitted and HEPA filter 4 Replace HEPA filter with the correct type see page 5 5 Replace cover filter and filter frame using the screws 6 Reconnect the mains supply and switch on 7 Check the airflow and adjust if necessary as detailed in the setup section 8 Filter should be DOP tested to verify filter integrity and results recorded REPLACEMENT FILTER P...

Страница 6: ...t i s h D e s i g n Manufacture Monmouth Scientific Limited UK Headquarters Units 5 6 Kilnside East Quay Bridgwater Somerset TA6 4DB 44 0 1278 458090 www monmouthscientific co uk info monmouthscientific co uk ...

Отзывы: