Immunity test /
standard
IEC 60601 - testing
level
Compliance level
Electromagnetic
environment guidelines
Surges
IEC 61000-4-5
±1 kV differential mode
voltage
±2 kV common mode
±1 kV differential mode
voltage
±2 kV common mode
The quality of the supply
voltage should be
comparable to that for a
typical shop or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
IEC 61000-4-11
< 5 % U
T
(> 95 % dip in U
T
)
for 1/2 period
40 % UT
(60 % dip in U
T
)
for 5 periods
70 % U
T
(30 % dip in U
T
)
for 25 periods
< 5 % U
T
(> 95 % dip in UT)
for 5 seconds
< 5 % U
T
(> 95 % dip in U
T
)
for 1/2 period
40 % UT
(60 % dip in U
T
)
for 5 periods
70 % U
T
(30 % dip in U
T
)
for 25 periods
< 5 % U
T
(> 95 % dip in UT)
for 5 seconds
The quality of the supply
voltage should be
comparable to that for a
typical shop or hospital
environment. We
recommend an uninterrupted
power supply or battery for
operators of the product
demanding continuous
function even during an
interrupted power supply.
Magnetic field in
power supply
frequency (50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Magnetic fields of the supply
frequency should conform
with the typical values found
in commercial or hospital
environments.
Note: U
T
is the alternating mains voltage prior to the application of the testing level.
The
Liposat
®
power
meets all test levels according to IEC60601-1-2 Edition 4
(Tables 4 to 9).
The requirements for use in aviation, transportation and military fields have not been taken
into account as they have not been tested.
Portable RF communications equipment (radio devices) (including their
accessories such as antenna cables and external antennas) should not be
used closer than 30 cm (or 12 inches) from the parts and cables of the
Liposat
®
power
indicated by the manufacturer. Non-observance may result
in a reduction of the device's performance.
Operation of the
Liposat
®
power
with additional accessories such as
transducers or cables, which are not defined for the intended use with the
device, may result in increased electromagnetic emissions, reduced
immunity to interference or faulty operation.