Docon SealM
Docon Seal
Remedying of errors
Page 39 of 68
Errors
Solution
The Docon SealM will not
charge.
•
Check the connection to the charger.
The displays on the charger and Do-
con SealM must light up or flash or-
ange.
•
If the displays fail to light up or flash:
Check the charger and socket or use
a different charger.
•
Recharge the Docon SealM.
If the energy-saving mode
will not switch on or off
•
Make sure that the battery is fully
charged (Docon SealM).
•
Make sure that the Docon SealM is
disconnected from the charger once
the battery is fully charged.
•
Make sure that the device is con-
nected to the mains (Docon Seal
only).
Error due to moisture in the
plug connection.
Remove all plugs from the device and the
seal handle and allow the connections to
dry.
Damaged, illegible inscrip-
tions or labels
Please contact the
manufacturer ‘s service
centre or your local distributor.