Molift Quick Raiser 2+ Скачать руководство пользователя страница 11

Page 11 of 24

Inspection Diagram after Service and Repair

MOLIFT Smart 150

Service:

  Checked service lamp if service is necessary:  YES (yellow or red light)  /  NO (green light)

Use Molift Servicetool and read the lifters data - check number of lifts left before service: _____________________

  Service is completed : - Worn or damaged parts, or malfunctioning parts, are replaced. See description for parts 

replacement. 

  Service light was reset using Molift Servicetool. N.B.: The service light must ONLY be reset after completed service!

  Perform safety inspection after service.

Safety Inspection

  The lift has been visually checked for damages, malfunction and deformation, and it is in order. Checked that lifting 

column is properly fastened to chassis, and that the lifter has no loose parts.

  Checked that mounting of column works correct. With legs in outer position (working position), it is possible to run 

the column up and down easily, and the column falls in place in the base. With lifter in transport position it is not 

possible to leave the column in upright position (it falls down when pushing slightly)

  Checked that remote control works and that wire and socket for remote control are undamaged. 

  Checked that the lifter functions correctly and runs without load all the way up and down without abnormal noise. 

Checked that end stop works. Repeated test with a load of approx. 100 kg. 

  The suspension and lifting arm are checked visually for damage, play and deformations. The suspension shall be 

turned 360° with some friction. If no defects or malfunctions are discovered, the suspension shall be tested with 

maximum load (150 kg). Run the lift up and down 2 - 3 times and turn the suspension 360° around 2 - 3 times with 

maximum load. Perform a new visual control; damage, play and deformations as described above. Any damaged 

parts must be repaired or replaced and the test is performed once more until the lift performs correctly.

  Checked that the leg spreading mechanism functions properly. Tested with maximum load. Verifi ed that the legs do 

not move beyond inner and outer limits (working position) when operated with load. The legs does not move inside 

parallel positions (or working position) with column mounted.

  Checked that emergency stop and lowering (electric and manual) are working correctly.

  Service light is green and batteries are suffi  ciently charged.

  Marked lift with safety sticker with month/year and certifi cate number for completed safety inspection. Inspecting 

person must sign safety sticker and this diagram.

Inspection, repair and service should be done by a person certifi ed by, or on behalf of Molift Group AS. 
Always complete below:

 

Client:________________________________ Serial number of lifter:_______________________________

 

Date: ________________________________  Signature:_________________________________________

Содержание Quick Raiser 2+

Страница 1: ...a patientlifter from T E C H N I C A L M A N U A L TM04301 Molift Quick Raiser 2 09 English...

Страница 2: ...ogue 12 Molift Quick Raiser 1 09 12 Main Components 13 Appendix A 14 Chassis w legs 15 Appendix B 16 Lifting column 17 Handle QR QR2 QR2 17 Appendix C 18 Cover with electronics QR2 19 Appendix D 20 Ba...

Страница 3: ...detecting the defect Remember to quote the serial number Consumable parts are not subject to the warranty All technical information data and instructions for operation contained in these operating in...

Страница 4: ...tube and steel plate frame with a polyester seal including legs The side housings are constructed in vacuum moul ded ABS The chassis consists of Motor with mounting plate and coupling Motor control ca...

Страница 5: ...icated and requires no further lubrication It is protected from dirt by two column wiper brushes Battery holder and batteries The battery pack is placed in a special alumi nium holder on the lifting c...

Страница 6: ...service immediately Service light codes Service light Mode No light Power saving Green Ready for use Yellow Book service Red Perform service Red sound Perform service immediately Flashing yellow Over...

Страница 7: ...out from the chassis Lifting motor fitting You will need Hex key 5mm Loctite 243 Procedure Follow the above procedure in the reverse sequence Put Loctite 243 on the motor s three mounting screws befo...

Страница 8: ...s Use a nozzle on the hot air gun so as to limit the heated area if possible Remove the wheel by unscrewing the 8x50 screw with the Umbraco key Fit the new wheel by follo wing these procedures in reve...

Страница 9: ...in the foot of the column on the chassis by unscrewing the two screws holding the unit Pull it out carefully so that you will be able to disconnect the wires with the soldering iron Disconnect the ol...

Страница 10: ...pair or service Logging Use the table below for logging of repair and work done on the lifter Give a brief description of the work carried out so that it is possible to see what has been done over a p...

Страница 11: ...hat end stop works Repeated test with a load of approx 100 kg The suspension and lifting arm are checked visually for damage play and deformations The suspension shall be turned 360 with some friction...

Страница 12: ...uick Raiser 2 see app B 2 1 0412013 Column Molift Quick Raiser 2 see app C 3 0455011 Knee Support large QR2 QR2 0455012 Knee support small QR2 QR2 optional 4 1 0541000 Battery 5 1 1340100 Battery Char...

Страница 13: ...Page 13 of 24 Main Components Appendix 1 5 2 4 3...

Страница 14: ...ctronics QR2 21 1 0421501 Leg Rods Assembly 22 2 0421108 Spacer Nylon 8 4x 12x20 23 2 0421109 Spacer Nylon 8 4x 12x25 24 2 3500211 Clewis Pin Headed HCB 10 X 35 25 2 3500218 Pin Split DIN 94 2 5x16 26...

Страница 15: ...8 03 2010 T B MHL D Update of pos no 12 02 2009 T B MHL C Added pos 39 29 01 2009 TES MHL B Uppdate of list 21 01 2009 T B Cable on actuator to be cut 100 millimetre long Assemble cable clips Blue fem...

Страница 16: ...310306 Label Art nr Id nr brady 17 1 1310307 Decal kontrollert brady 18 0410117 Cable Gland RG2 19 0410118 Cable Gland 9x11x16 20 1 0420150 Connection Bracket 21 1 0220102 Screw DIN 912 M10X14 ELZN 22...

Страница 17: ...15 14 13 12 11 10 9 8 7 6 5 4 3 2 31 Pos Qt Molift Quick Raiser Drawing name 17 12 2008 Drawing number Coloumn assy QR2 2008 A2 08 174 1 5 T B Projection Scale Date Designed Drawn Size Appendix B Geom...

Страница 18: ...8 6 0421120 Rivet Plastic 5x4 7 Black 9 0 Screw part of 0421120 pos 8 10 1 0220307 Label Emergency Stop 11 1 0421121 Label Panel QR 2008 12 1 0421122 Extension Emergency Stop QR 13 1 0421123 Cable Ex...

Страница 19: ...to come out in rear of cover Put in a lope and into the two camps on side 13 17 20 19 18 B B 17 Mount in front wall in front of track for PCB 21 9 8 9 8 22 C A A B SECTION A A SCALE 1 2 C Added rubber...

Страница 20: ...e holder 2 6 0540501 Screw Wallboard Elzn 3 1 0510205 Lower part plastic 4 1 0510510 Upper cover mounted 5 6 7 8 1 0540505 Rubber bushing 9 2 0540120 Screw DIN965A Elzn 10 2 0540123 Nut w threads bras...

Страница 21: ...e specisl tool to mount 15 Label with art no on back side of unit Assembly instruction 08 P B B 11 50 Rev Description Date Sign Approved C Pos 4 5 6 and 7 replaced by new pos 4 07 10 2008 TES MHL B Ch...

Страница 22: ...01 2009 T B MHL Rev Description Date Sign Approved Designed T Leg Drawing n Mol Appendix E Pos Qty Art no Description Fig 1 0455011 KNEE SUPPORT COMPLETE 1 1 0455101 Knee support bracket long model 2...

Страница 23: ...lack Red Male Male Female Female Female Connection point chassis column Red Red Black Black Male Male Female Female NiMh 14 4 V Red Green Black Red 13 PIN DIN cable Hand set Chassis Rev Description Da...

Страница 24: ...land Meditec Medical Ltd Unit 28 Whitestown Drive Whitestown Industrial Estate Tallaght Dublin 24 Tel 01 4624045 www meditecmedical ie Israel Mediscan Systems Ltd P B Box 2195 6 Hamanov St Rehovot 763...

Отзывы: