Page 4 of 20
BM06101 Molift MOVER 300 English
Rev E 24/09/2013
General safety precautions
General
Before putting a Molift Mover 300 into operation,
the operator must ensure that the lifter and acces-
sories are in good working order, especially with
regards to safety.
A Molift Mover 300 shall only be used when all
safety devices and equipment is in place and fully
operational.
The lifter must be taken out of service immediate-
ly, or not returned to service, if faults and/or irreg-
ularities that may have influence on the safety are
discovered.
The lifter has an expected lifetime of 10 years
or 30 000 lifts with max load (SWL), when all
recommended service is performed.
Injury to personnel or equipment may occur if
Molift Mover 300 is used in an incorrect manner
and if i.e.:
• covers are removed by unauthorised personnel
• used incorrectly
• insufficient maintenance
• load exceeds maximum limit. The maximum
load stated in this manual and on the lifter is
the maximum load.
• unauthorised repair of electrical devices or ca-
bles.
Lift and transfer
The Molift Mover 300 has a Safe Working
Load of 300 kg/660 lbs, meaning that it is
tested and approved for lifting of patients
up to 300 kg/660 lbs. Attempts to lift patients
above SWL may damage the lifter and lead to
personal injury.
The patient lifter may tip over if used incor-
rectly! Ignorance of warnings and instruc-
tions may lead to personnel injury. Read in-
structions carefully before attempting to lift
anyone.
Pay close attention to your actions. Ensure
that all four loops of the sling are securely fas-
tened to avoid the user slipping or falling.
Use the push handle to manoeuvre the lift, do
not attempt to manoeuvre the lift by pushing
on the lifting arm, actuator or arm.
Torn, cut, frayed or broken slings can fail, re-
sulting in serious injury to the user. Use slings
in good condition only. Destroy and discard
old, unusable slings.
Molift slings shall only be used to lift persons.
Never use sling to lift and/or move objects of
any kind.
Molift Mover 300 should not be run con-
stantly for more than 2 minutes (with maxi-
mum load), and rest for minimum 18 min-
utes.
If the lifting movement does not start
when the corresponding button is pressed,
please consult the fault finding table in the
maintenance chapter to locate the fault.
Several factors have to be considered during
movement of the lifter. Stand to the side of the lift-
er when lifting. Ensure that arms and legs are clear
of the lift, chair, bed etc. try to maintain eye con-
tact with the user - this will help the user feel safe
and comfortable.
Medical electrical equipment requires spe-
cial precautions regarding electromagnet-
ic compatibility (EMC). Portable or mobile
radio communication equipment may af-
fect the medical electrical equipment, and
should be kept minimum 25 cm (10 inches)
from the lifters electronics