VersaMax, SpectraMax 340PC384, 190, Plus 384 Microplate Readers User Guide
46
0112-0126 B
Item
96-wells
Endpoint standard read
9 seconds single wavelength
19 seconds dual wavelength 425 & 650 nm
Kinetic read intervals
9 seconds minimum interval between reads single wavelength
1 column, 2-second minimum interval between reads
Single wavelength
Measurement Time (calibration off) SpectraMax 190
Item
Description
Plate: normal scan
33*K nm/min (8-well strip)
Plate: speed scan
135*K nm/min (8-well strip)
Scan Speed (*K = wavelength interval) SpectraMax 190
Item
Description
Read chamber
Isothermal when temperature regulation is not enabled
Range
4°C above ambient to 45°C when temperature regulation enabled.
The ambient temperature must be >20°C to achieve temperature
regulation at 45°C.
Resolution
± 0.1°C
Accuracy
± 1.0°C for plate chamber
Temperature uniformity
at equilibrium
± 0.5°C at 37°C
Chamber warm-up time
15–30 minutes (measured on air) after initiation of temperature
regulation
Temperature regulation
4 sensors
Drift
± 0.2°C (regulated)
Temperature regulation
diagnostics
Temperature regulation system is continuously monitored and updated
Evaporation
Plate lid required to minimize evaporative cooling
Recommended plate
Flat-bottom plates with isolated wells and lid
Temperature Regulation SpectraMax 190