Setting Up the SpectraMax® MiniMax™ 300 Imaging Cytometer
5024111 D
15
Removing Transport Locks
CAUTION!
The instrument can be damaged if the you do not remove the transport
locks before you power on the instrument.
There are transport locks on the detection cartridge drawer and the plate drawer to protect
the instrument from damage during shipping. You must remove the transport locks before
you power on the instrument.
Note:
If you install the SpectraMax MiniMax 300 Imaging Cytometer, leave the
transport locks in place until after you attach the cytometer to the microplate reader.
See the installation guide or user guide.
Illustration
Part Number
Description
YW 000 006
2.0 mm hex key
Required Tool
CAUTION!
Do not touch or loosen screws or parts other than those specifically
designated in the instructions. Doing so could cause misalignment and possibly void
the warranty.
CAUTION!
The front cover is held onto the front of the instrument by powerful
magnets. Keep magnetic storage devices or strips, such as hard drives, key cards, and
credit cards, away from the instrument covers.
To remove the transport locks:
1. Firmly pull on the front cover from the bottom to remove it. The front cover is held in
place by magnets.