Setting Up the SpectraMax® i3x Multi-Mode Detection Platform

5025026 C

9

7. Gently pull the yellow tab that protrudes from the plate chamber door to open the

door. Hold the door (5) open while you remove the transport lock.

Note:

Do not to tear the yellow tab. It remains attached to the transport lock

to make it easier to open the plate chamber door.

Item

Description

Screw #1 fastens the lock to the internal frame of the instrument

Screw #2 fastens the lock to the plate drawer

Screw #3 fastens the lock to the plate drawer

Plate drawer

Plate chamber door in open position

Plate drawer transport lock

8. Use the 2.0 mm hex key to loosen screw #1 in the upper-left corner of the transport

lock until the lock disconnects from the instrument frame. The screw has a retaining
washer to prevent removal from the lock.

Tip:

After you loosen screw #1, pull the plate drawer slightly out of the

instrument to hold the chamber door open.

9. Loosen screws #2 and #3 until the lock comes free of the plate drawer and you can

remove the lock from the instrument. The screws have retaining washers that
prevent removal from the lock. Store the transport lock in the accessories tool box.

10. Push the plate drawer back inside the instrument and close the chamber door.
11. Save the carton, foam inserts, accessories tool box, and transport locks for future

shipments.

Содержание SpectraMax i3x

Страница 1: ...5025026 C October 2018 SpectraMax i3x Multi Mode Detection Platform Installation Guide...

Страница 2: ...right and or license to use or permit others to use such manufacturers and or their product names as trademarks Each product is shipped with documentation stating specifications and other technical in...

Страница 3: ...t up the instrument Please be aware that some updates to the SoftMax Pro Software require a purchase Contact Molecular Devices before you update the software To download the latest version of the soft...

Страница 4: ...the contents of the package see the enclosed packing list Note Detection cartridges are shipped separately The SoftMax Pro Software installation places a copy of the available microplate reader user g...

Страница 5: ...mum of two people to lift the instrument Note Retain the shipping box and all packaging materials for future transport needs Do not use tools that can damage the packaging or the instrument CAUTION Wh...

Страница 6: ...Tip Have a second person hold the box in place while you slide the instrument out on the cardboard 4 Remove the accessories tool box 5 Remove the foam packaging and the plastic bag 6 With one person o...

Страница 7: ...ocks in place until after you attach the cytometer to the microplate reader See the installation guide or user guide Illustration Part Number Description YW 000 006 2 0 mm hex key Required Tool CAUTIO...

Страница 8: ...r 2 Detection cartridge drawer transport lock 3 Slide the detection cartridge drawer forward until it is outside of the instrument 4 Lower the detection cartridge drawer transport lock to remove it fr...

Страница 9: ...ition 6 Plate drawer transport lock 8 Use the 2 0 mm hex key to loosen screw 1 in the upper left corner of the transport lock until the lock disconnects from the instrument frame The screw has a retai...

Страница 10: ...8 foot 4400 0002 or 4400 0036 Power cord 1 meter 3 3 foot Item Description 1 USB port 2 Gas inlet quick connect fitting 3 Power port 4 Fuse carrier 5 Power switch To connect the instrument cables 1 M...

Страница 11: ...e tubing outside diameter 4 0 mm Not applicable Flat head screwdriver S MS 135 100 Gas inlet unlock tool not provided with the instrument Item Description 1 USB port 2 Gas inlet quick connect fitting...

Страница 12: ...djust the gas supply input pressure Note For the maximum permitted air supply pressure for the gas inlet and polyurethane tubing specifications see the user guide for the SpectraMax i3x To disconnect...

Страница 13: ...cm 7 9 in to 11 8 in gap between the rear of the instrument and the wall To turn the instrument on or off press the power switch to place the rocker in the on or off position Item Description 1 Power...

Страница 14: ...ty of Molecular Devices LLC or their respective owners Specifications subject to change without notice Patents www moleculardevices com productpatents FOR RESEARCH USE ONLY NOT FOR USE IN DIAGNOSTIC P...

Отзывы: