Operation
20
5001410 B
10.
On the Image tab, press
Ctrl+N
to create a New Settings file.
A scan area is created that matches the size of each slide. You
can reduce the amount of time for the scan by reducing the scan
area size.
11.
Click
Scan Area Properties.
12.
Type the following values:
Left
: 3500
Top
: 5000
Width
: 15000
Height
: 3000
13.
Click
OK
.
14.
On the
Batch Scan
tab, click
Scan Batch
.
The batch scan begins and all four slides are loaded, scanned,
and unloaded, based on the defined settings.
The GenePix Pro Software automatically logs hardware
performance into the Hardware Diagnostics Report every two
hours during operation.
Manual Scanning
In Manual scan mode, the GenePix Pro Software operates the scanner
only, in a conventional way. You can open and close the motorized
scanner door, perform Preview Scans, Data Scans, adjust scanner
settings from the Hardware Settings dialog box, and stop and start
scans.
Any Global file options (Save locations, File name) set during a batch
scan apply when performing a manual scan.
Follow the instructions carefully when inserting a slide in the scanner
slide holder. The slide holder is a precision component designed to
ensure proper focusing and field uniformity. Improper handling may
damage the slide holder and affect imaging performance. In Manual
scan mode, the GenePix SL50 Slide Loader is inactive and will not
move.
GenePix SL50.book Page 20 Tuesday, February 1, 2011 12:20 PM