Installation and Operation
14
5000450 D
To unpack the scanner and position it on the bench
1.
Open the crate and remove any foam blocks used to protect the
scanner from excessive shock and vibration during shipping.
CAUTION!
Do not touch or loosen any screws or parts other than those
specifically described in the instructions. Doing so can cause
misalignment and voids the scanner warranty.
2.
Lift and then place the scanner on a flat level surface, away from
direct sunlight, dust, drafts, vibration, and moisture. Carefully
move the scanner to its final position.
3.
Plug the USB cable into the USB type B port on the back of the
scanner, and the USB type A port on the control computer.
.
Note:
Do not tilt or slide the scanner when transferring it from
the crate to its final position on a bench or desk.
Tip!
For proper ventilation and cooling, leave at least six inches
of space between the back of the scanner and the nearest
object or surface.
Figure 2-1
Connections on back of the scanner
Item Name
1
Round port for main power supply cable
2
USB type B port
2
1