FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
27
Consumables and Accessories on page 283
Cell Suspension
The Cell Suspension option consists of two components:
•
The Cell Reservoir installable in Position 4 (Source 3) in the 5
position stage.
•
The Cell Suspension module located externally on the right side
of the instrument.
The Cell Reservoir (see
) is user installable. The Cell
) consists of a shelf with a magnetic
motor mounted underneath it, a cell flask with a magnetic stirrer, up to
four fluid bottles for automated cleaning, and a removable cover for
keeping cells in a dark environment.
demonstrates how the system is connected and shows all
the possible combinations for protocol development.
Figure 2-5
Cell Reservoir.
The Cell Reservoir, which the user places in a source plate location, is
filled from any of the bottles in the external Cell Suspension module by
a pump with adjustable speed and direction.
The Cell Reservoir is a special plate type that has one fluid line used for
both input and output, and an electronic plate ID that is part of the
electrical/fluidic connector which identifies the reservoir to the system.
In the case of an overflow, an overflow trough catches excess fluid and
pipes it to the tip washer overflow trough, which directs it to the
overflow bottle. The reservoir is also autoclavable for cleaning
purposes.
The spinner flask contains a stirrer, which is driven by a magnetically
coupled motor mounted in the lower part of the Cell Suspension
module. The Cell Suspension stir speed can be set in a protocol or with
manual instrument controls. Stir speed of 5 equals approximately 1
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...