C77–Parameter is out of range. Range is …
Washcellreservoir<TAB>Fluid Type, Fill Speed, Drain
Destination, Drain Speed, Wash Cycles, Hold Time,
Volume Level<CR>
Command Description
This command is the equivalent of selecting
Instrument/
Manual
Operation
/
Wash Cell Reservoir
from the main menu of the
ScreenWorks Software. This selection will wash the Cell Reservoir with
the parameters provided.
Command Parameters
There are seven required parameters for this command. All seven
parameters must be supplied for this command to properly execute.
The following table lists the parameters, their purpose, and the
appropriate range for each parameter.
Table A-2
Washcellreservoir<TAB>Fluid Type, Fill Speed, Drain
Destination, Drain Speed, Wash Cycles, Hold Time, Volume Level<CR>
command parameters
Parameter
Type
Description
Acceptable values
or range
Fluid type
string
Indicates the fluid bottle
to be used.
Fluid 1, fluid 2, fluid 3,
fluid 4, cell flask.
Fill speed
integer
Speed to run the pumps
which fill the Cell
Reservoir. Range from
1–10 where 1 is very
slow, and 10 is very fast.
1–10
Drain
destination
string
Indicate the bottle to
drain fluid to.
Fluid 1, fluid 2, fluid 3,
fluid 4, cell flask, waste
a, waste b.
Drain speed
integer
Speed to run the pumps
which drain the Cell
Reservoir. Range from
1–10 where 1 is very
slow, and 10 is very fast.
1–10
Wash cycles
integer
Number of wash
reservoir fill/drain cycles
to complete.
1,2,3,4,5
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...