Robotic Integration
252
0112-0109 H
Command Parameters
There is one parameter, folder, for this command.
The folder parameter identifies the folder for which a list of all *.fmp
type files will be returned. The folder must be complete and in the
correct Windows format, including the drive identifier and a closing
backslash(\).
Example Command Line
findprotocols<TAB>c:\screenworks\protocols\<CR>
Example Response
[myprotocol1.fmp,myprotocol2.fmp,myprotocol3.fmp]<CR> or c45,
error string <CR>
Suggested Steps in Command Usage
•
Send
findprotocol
command at any time.
•
Response will be list of protocols separated by comma or
Condition code related to command, that confirming command
could not execute.
Command Specific Error Responses
•
Folder does not existCondition code is C45
•
Invalid folder nameCondition code is C46
•
Current document is not valid protocolCondition code is C50
•
No protocols found in dirCondition code is C56
Runexperiment<TAB>[Data File Name]<CR>
Command Description
This command will run the currently open protocol, if it is valid for the
current instrument configuration.
Running any protocol will create a data file and shift the software focus
to this data file for the duration of the experiment. When running in
Remote mode, the data file will be closed automatically when
experiment is done and focus will automatically return to the protocol
file.
Command Parameters
There is one optional parameter, data file name, for this command.
If a data file name parameter is used with this command, the data file
created by running the current protocol will be named data file name.
The filename will be appended to the
Data
folder path specified in the
protocol
Settings
screen in the ScreenWorks Software. Never specify
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...