Troubleshooting
218
0112-0109 H
Troubleshooting the Cell Reservoir
Troubleshooting Data
Table 9-8
Troubleshooting the Cell Reservoir
Symptom or Error
Message
Possible Causes
Solutions
255 “CellReservoir: Reservoir
not detected. Please install it at
Source 3 position and select
Refresh Configuration
from
Instrument
menu bar.”
Instrument cannot
detect the cell
reservoir.
Check to make sure
the Cell Reservoir is
installed.
Check the connection
of the reservoir
sensor inside the
instrument.
256 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 1”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
1.
Check to make sure
that Fluid 1 bottle is
not empty.
257 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 2”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
2.
Check to make sure
that Fluid 2 bottle is
not empty.
258 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 3”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
3.
Check to make sure
that Fluid 3 bottle is
not empty.
259 “CellReservoir: prime sensor
dry during prime/fill. Please
check Fluid Bottle 4”.
Fluid sensor is not
detecting fluid when
pumping from Fluid
4.
Check to make sure
that Fluid 4 bottle is
not empty.
Table 9-9
Troubleshooting Data
Symptom or Error
Message
Possible Causes
Solutions
Negative control wells
show a response.
Wash buffer
components are
different from the
compound buffer.
Make sure wash buffer
and compound buffer are
the same. Check for a
DMSO stimulation, if
present.
Statistical results do not
conform to the expected
results.
Data analysis has not
been optimally set up.
Refer to
.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...