Troubleshooting
210
0112-0109 H
Troubleshooting the Optics
“Done with plate after
dispense” is not present.
TETRAcycler is not
present on
instrument.
Must install TETRAcycler on
instrument for command to
be available.
Dialog does not
appear because you
are dispensing to
the Read plate
position.
Read plate is automatically
changed at the end of an
experiment when in
Remote Mode
.
“Mix Fluid After Dispense”
is not present.
Command is not
available when
“Single aspirate –
Multiple dispense”
fluid transfer type is
selected.
Select either
Single
aspirate – Single
dispense
or
Multiple
aspirate – Single
dispense
fluid transfer
types.
Table 9-4
Troubleshooting the Pipettor (cont’d)
Symptom or Error
Message
Possible Causes
Solutions
Table 9-5
Troubleshooting the Optics
Symptom or Error
Message
Possible Causes
Solutions
240: “Attempting to use a
LED module that is not
installed. Please install
the desired LED banks.”
Specified LED module
in protocol is not
installed.
Install the appropriate
LED banks.
Select the appropriate
LED module that is
already installed on the
instrument.
241: “Attempting to use
an emission filter that is
not installed. Please
install the desired
emission filter.”
Specified emission filter
in the protocol is not
installed.
Install the appropriate
emission filter.
Select the appropriate
emission filter that is
already installed on the
instrument.
210: “Unable to read
plate. Calibration not
valid for
emission/excitation pair.”
A calibration file is not
available for the
LED/emission filter
combination selected.
Calibrate instrument for
desired LED/emission
filter pair.
Содержание FLIPR Tetra
Страница 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Страница 12: ...Contents 12 0112 0109 H...
Страница 16: ...System Overview 16 0112 0109 H...
Страница 40: ...System Hardware Features 40 0112 0109 H...
Страница 148: ...Exchanging Hardware 148 0112 0109 H...
Страница 156: ...Calibration and Signal Test 156 0112 0109 H...
Страница 196: ...Running an Experiment 196 0112 0109 H...
Страница 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Страница 282: ...Data Processing Algorithms 282 0112 0109 H...
Страница 294: ...Consumables and Accessories 294 0112 0109 H...
Страница 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Страница 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...