Opening a Drawer Manually
Under some circumstances it might be necessary to open an instrument
drawer manually.
1.
If the power is not already off (as can occur with a power failure,
or if you are already troubleshooting the instrument with the
covers open), turn off the power. Disconnect the power cord
from the instrument, and the computer cables from the
computer.
2.
Using your fingernail, or a narrow blade, open the drawer door
to its horizontal position and pull out the carriage slowly, and as
gently as possible.
WARNING! The power must be off for this procedure. If the
power is not disconnected, and a jammed carriage or drawer is
suddenly freed, the instrument could suddenly move, resulting
in operator injury.
CAUTION!
Never force a carriage out of the instrument. If it
does not come out reasonably easily, contact Molecular Devices
Technical Support. See