FlexStation 3 Benchtop Multi-Mode Microplate Reader User Guide
114
0112-0127 B
Obtaining Support
Molecular Devices is a leading worldwide manufacturer and distributor
of analytical instrumentation. We are committed to the quality of our
products and to fully supporting our customers with the highest possible
level of technical service.
Our support web site,
www.moleculardevices.com/support
, has a link to
the Knowledge base with technical notes, software upgrades, and other
resources. If you do not find the answers you are seeking, follow the links
to the Technical Support Service Request Form to send an email message
to a pool of technical support representatives.
You can contact your local representative or contact Molecular Devices
Technical Support by telephone at 800-635-5577 (U.S. only) or
+1 408-747-1700. In Europe call +44 (0) 118 944 8000. Please have the
system ID number, system serial number, software version number, and
the system owner’s name available when you call.
Moving the Instrument
If you need to relocate the FlexStation 3 instrument, follow these steps.
1.
Ensure that the new location meets the proper specifications as
described in
.
2.
Remove any microplates, adapters, reservoirs, and tips from all
three drawers, and then close the drawers.
3.
Be sure reading chamber screws are still in place. Contact
Technical Support if reading chamber screws have been
removed.
4.
Be sure all covers are closed and securely fastened.
WARNING! The instrument weighs approximately 50 pounds and
should be lifted with care. To prevent injury, use at least two people to
lift the instrument.
CAUTION!
Do not attempt to assemble or disassemble the
instrument with the pipette tips or compound and read plates in
place. Spillage or damage to the pipette tips, plates, or the
instrument can occur.
FlexStation_manual.book Page 114 Tuesday, June 25, 2013 10:22 AM