3. Operating Instructions
Figure 3.28.
Envelope Plot with Fill Mode, showing the range of impedance values at
the cursor position (51.5–185.4 ohms, red circle)
3. Select the
Math
option from the
Scan
menu and choose
Set Base
. The selected trace will
turn purple and is now a base trace for further
Math
functions.
4. Press the
SELECT
button again until the second trace is selected.
5. Select
Difference
from the
Math
menu. A new trace is created that displays the difference
between the base trace and the selected trace. See Figure 3.29, which shows 3.43 ohm
excess impedance due to a slightly loosened BNC barrel adapter. The bottom bold trace is
created by subtracting the top (live) trace from the middle (scanned) trace and adjusting
the vertical scale to emphasize trace detail.
Figure 3.29.
Difference trace. Live (base trace, purple, top), comparison scan (red,
middle), difference (blue, bottom) traces show 3.43 ohm excess impedance (red circle) on
the live trace due to a loosened BNC connector.
46
CT100B TDR Cable Analyzers Operator’s Manual