D.5 Auxilliary measurements
67
implemented in current models.
DAC
dac,v
Set voltage on
SMA
connector,
±
2
.
5 V 12-bit.
v can be an integer: 0
≤
v
≤
4095 corresponding to
−
2
.
5
..
+ 2
.
5 V
or a floating-point number
−
2
.
5
≤
v
≤
+2
.
5 but in that case must include a
decimal point. For example
dac,2
will be interpreted as an integer (equiv-
alent to approximately
−
2 V) but
dac,2.
and
dac,2.0
will be interpreted
as floating-point (2
.
0 V).
POWER
pow
Report power supply voltages.
TEMP
temp,chan
Report temperature.
chan=1,2
for main unit and imaging sensor respec-
tively.
PRESSURE
press
Report pressure in hPa.
press,Pactual
Define actual pressure (hPa). The value will be used to recalibrate the
pressure sensor.
PRAW
praw
Report raw pressure in hPa. Pressure correction stored in eeprom will be
programmed into pressure sensor.
WDG
wdg,wdtime
Set the watchdog time, in seconds.
wdtime = 3..16
(default 3 s).
The watchdog timer automatically resets the device if the device firmware
freezes.
For example, if the watchdog time is set for 1 second, and a
2 second exposure is initiated, the watchdog timer will reset the device
halfway through the exposure. This facility allows automatic restart after
interruptions or if there are software or firmware bugs.
ONTIME
ontime
Time in seconds since last reset or power-on.
Содержание MWM
Страница 1: ...MWM wavemeter Revision 3 19 mogwave 1 4 15 Firmware 0 6 9 ...
Страница 4: ...ii ...
Страница 7: ...Contents v G Ugrading firmware 83 References 84 ...
Страница 8: ...vi Contents ...
Страница 54: ...46 Chapter 6 Calibration ...
Страница 58: ...50 Appendix A Specifications ...
Страница 66: ...58 Appendix B Communications ...
Страница 70: ...62 Appendix C Programming ...
Страница 86: ...78 Appendix E mogwave configuration ...
Страница 92: ......
Страница 94: ...86 ...
Страница 95: ......