20
Chapter 4. Operation
Figure 4.7:
The recalibration window of the host software allows correction
of the device calibration using a known reference.
A standard reference
can be selected from the dropdown box, or a custom reference frequency
can be entered.
tion can also be applied programmatically using the
MEAS,CORRECT
command.
Note that when adjusting the calibration, it is recommended to use
a well-known reference (e.g. atomic transition) with an appropriate
amount of averaging.
The calibration can also be reverted to the factory-provided cali-
bration by ticking the appropriate box in the software interface, or
using the command
MEAS,CORRECT,RESET
.