3.3 Optimisation
23
Figure 3.2:
Scanning the laser with P-only feedback on the fast output
while scanning the slow output causes the error signal (orange) to become
extended when the sign and gain are correct (right). In a PDH application,
the cavity transmission (blue) will also become extended.
•
Set
SLOW GAIN
to mid-range and
SLOW INT
to
∼
100 Hz.
•
Set
SLOW
mode to “
LOCK
”. If the servo unlocks immediately,
try inverting the
SLOW SIGN
. It may also be necessary to make
small adjustments to the
ERR OFFSET
and associated trimpot.
•
Adjust
SLOW GAIN
and
SLOW INT
for improved lock stability.
•
Some applications may benefit by increasing
FAST DIFF
to im-
prove loop response.
3.3
Optimisation
The purpose of the servo is to lock the laser to the zero-crossing of
the error signal, which ideally would be identically zero when loc-
ked. Noise in the error signal is therefore a measure of lock quality.
Spectrum analysis of the error signal is a powerful tool for under-
standing and optimising the feedback. RF spectrum analysers can
be used but are comparatively expensive and have limited dynamic
range. A good sound card (24-bit 192 kHz, e.g. Lynx L22) provi-
des noise analysis up to a Fourier frequency of 96 kHz with 140 dB
dynamic range.