Product Warnings

Patient/Staff Safety

Patients, especially children must not be left unattended on or with a stretcher at any time.
Care must be taken in the operation of all moving parts and accessories to minimise risk of injury to 
staff or patients.
This stretcher is not suitable for use as a ward bed. Use for more than a few hours could contribute to 
patient discomfort and increase the potential for injury, including development of pressure injury.
The stretcher top must be in the lowest position when patients are getting on or off the stretcher, with 
access from either side only.
Maximum static patient load capacity on the backrest must not exceed 100kg distributed evenly or 
damage to locking struts and frame may occur.
Maximum static patient load capacity on the stretcher is 300kg and must not be exceeded. Weight 
must be distributed as evenly as possible on stretcher top. Any weight loading above 150kg must be 
wheeled on smooth, level flooring and attended by more than one staff member.
Only apply brakes when stretcher is stationary.
Brakes are to be fully applied during bed / stretcher transfers and attended by qualified clinical staff.

General Use

The backrest must be raised and lowered with care to prevent possible strain to the operator.
Check that all fingers, hands, limbs or x-ray cassette components are clear before lowering backrest.
Check that there are no obstructions above or below the top before adjusting the height.
The stretcher must never be left on sloping surfaces unattended, even with brakes fully engaged.
Traversing side on, across slopes is to be avoided under any circumstances.
On wet or slippery flooring do not lean or apply unnecessary force against stretcher, as it may slide: 
even with brakes fully applied.
This stretcher is specifically designed for indoor use in dry environments, with due care it may be used 
to transit outdoor areas however additional controls and precautions specific to the environment will 
be merited.
Clips, locks and retainers are supplied by the manufacturer for patient / staff safety and must not be 
removed.

Dropsides

Stretcher must never be raised or moved with dropsides down.
Patients on a stretcher with dropsides down, must always be attended by qualified clinical staff.
Dropside releases must be operated only by suitably qualified clinical staff in attendance.
Check that all fingers, hands and limbs are clear before lowering dropsides.
Always ensure dropsides are locked firmly in position after raising.
Side loading on any dropsides should not exceed 80kg or damage to the stretcher may occur.
Dropsides locked in mid position should not be subjected to any downward force exceeding 30kg static 
loading as this may result in damage to the stretcher or personal injury.

v-14092022 - Page 4

"Trusted products made for trusted professionals"

Содержание Contour Multi-X

Страница 1: ...Contour Multi X Electric User Guide...

Страница 2: ...and or treatment within the environment A range of models designs features and accessories are available for this customisable device These instructions for use are for powered electric battery stret...

Страница 3: ...Product Overview Overall Size 2135 x 815mm overall size 705mm narrow or 965mm wide top options Height Range 545 915mm SWL 300kg v 14092022 Page 3 Trusted products made for trusted professionals...

Страница 4: ...ssette components are clear before lowering backrest Check that there are no obstructions above or below the top before adjusting the height The stretcher must never be left on sloping surfaces unatte...

Страница 5: ...the cover using a LAUNDRY SAFE FELT TIPPED MARKER Electric Operation All electric functions must be operated only by suitably qualified clinical staff in attendance Stretcher top must not be lowered o...

Страница 6: ...dvice PATIENT SAFETY REQUIRES THAT YOU ALWAYS CHECK THE MATTRESS BEFORE USE A Modsel Protran Premia or Premia Forte mattress is specifically designed for use in conjunction with Modsel Contour stretch...

Страница 7: ...that a stretcher with corner steering is always pushed from the opposite end to which the steer castor is positioned however the end from which a stretcher is pushed from is subject to an organisation...

Страница 8: ...e lock tab and then lower the dropside Ensure the dropside is supported to control speed of lowering Always check that all fingers hands and limbs are clear before lowering dropsides When the dropside...

Страница 9: ...s the DOWNWARD facing arrows labeled TILT FOOT HEAD on the handset to lower either end of the stretcher top To raise and lower knee break Press the UPWARD facing arrow labeled KNEE on the handset to r...

Страница 10: ...sing the quick release Always check that all fingers hands limbs or x ray cassette components are clear before lowering backrest If the backrest quick release is activated lower the backrest completel...

Страница 11: ...ce the battery life Regular charging of the battery is recommended On hearing an audible low battery warning tone from the unit during use remove the battery for charging immediately To charge the bat...

Страница 12: ...be operated while powered via a standard 240V outlet There is a battery backup to allow stretcher operation whilst unplugged ALWAYS ENSURE THE POWER LEAD IS UNPLUGGED AND STOWED CORRECTLY BEFORE MOVI...

Страница 13: ...then either one at a time or with both hands upside down and using your forefingers depress the lock buttons and allow the accessory to drop in Ensure both locks have activated after installation Rem...

Страница 14: ...e been supplied separately Spot clean any visible soiling Wipe over all surfaces with a suitable disinfectant ensuring a suitable contact time Rinse or remove residues as merited Store indoors in a cl...

Страница 15: ...d wipe surface as clean as possible and sanitise using a suitable disinfectant If a zipper is present fully unzip the cover and remove mattress core TAKING CARE not to contaminate the foam core Store...

Страница 16: ...ccordance with our recommended Service Checklist An approved MODSEL Service Checklist can be downloaded online from this link TO ENSURE WARRANTY PERIODIC MAINTENANCE RECORDS MUST BE RETAINED Malfuncti...

Страница 17: ...have a manufacturing defect within 3 three months from the date of invoice The purchaser within 14 fourteen days of discovery of the defect or defects gives written notice to Modsel or their distribut...

Страница 18: ...4 56 Industrial Ave Maryborough QLD 4650 Australia Contact 61 1300 785 790 Class I Medical Device ARTG Included Product Model Serial Number Mattress Model Batch Number Date of Manufacture v 14092022 P...

Страница 19: ...Trusted products made for trusted professionals Notes v 14092022 Page 19 Trusted products made for trusted professionals...

Страница 20: ......

Отзывы: