Quick Setup
Quick Setup
Quick Setup
Quick Setup
CONNECTING THE AIR COMPRESSOR TO THE LIQUEFIER:
1. Connect Air Hose from Air Compressor to Liquefier. Hand tighten only.
2. Connect Air Compressor Data cable to Liquefier and Air Compressor.
3. Connect both power leads from Air Compressor and Liquefier to suitable wall
outlets.
4. Turn on Air Compressor and Liquefier.
5. Switch the device to ‘Run’ 11
SETTING THE CLOCK (IF NECESSARY):
The clock is factory set, and is displayed when the
unit powers up, followed by the day. Press ‘Reset ’13 to view the clock at any time.
See page 8 if the time or date are incorrect. The clock is used for automatic
maintenance, which takes place at 9PM every Saturday night.
TESTING THE AIR SUPPLY:
With the unit running, press and hold the ‘Test Air Pressure’
button 14 on the rear of the unit. Over a period of 30 seconds, the ‘Pressure’ 10
should rise to a minimum of 90psi. If it doesn’t reach 90psi, check the air hose
connections for leaks.
DEFROST:
Each time the unit starts up following installation, a power failure, or resetting of
the unit, a 90 minute defrost is commenced. This can be cancelled at any time by
pressing the ‘Defrost’ button 8 on the front panel. Subsequent defrosts can be
started if required by pressing and holding the ‘Defrost’ button for 2-3 seconds.
VIEWING SENSOR VALUES:
When the system is not defrosting pressing the ‘Mode’ button
6 will toggle the sensor value as indicated by the lights 1
COOLDOWN:
Once defrosting has completed, the refrigeration system will become active
and will begin to cool down the liquefier. This can be monitored by watching the
“Cryo Temp”.
See Viewing Sensor Values
. When the Temp has reached 77K, cool
down is complete and production will begin. This should take about 30 minutes.
LOCKING AND UNLOCKING:
The ‘Lock’ feature 9 is a security device to stop unwanted
dispensing. If the unit is unlocked and the Dewar is in place, dispensing is enabled
indicated by the ‘Ready’ light 2. To unlock the unit, first hold down the ‘Lock’ button
9. While holding the ‘Lock’ button, press and hold the ‘Mode’ button 6
simultaneously until the lock light goes out. To relock, simply hold the ‘Lock’ button 9
until the lock light comes on.
Note: When locked, the unit will produce internally and
stay in the Dewar; it will maintain a full level in the liquefier until the device is
unlocked. Unlocking can only take place in normal operation.