Quick Setup

Quick Setup

Quick Setup

Quick Setup    

 

CONNECTING THE AIR COMPRESSOR TO THE LIQUEFIER: 

1. Connect Air Hose from Air Compressor to Liquefier. Hand tighten only.  
2. Connect Air Compressor Data cable to Liquefier and Air Compressor. 
3.  Connect  both  power  leads  from  Air  Compressor  and  Liquefier  to  suitable  wall 

outlets. 

4. Turn on Air Compressor and Liquefier. 
5. Switch the device to ‘Run’ 11 

 

SETTING  THE  CLOCK  (IF  NECESSARY): 

The  clock  is  factory  set,  and  is  displayed  when  the 

unit powers up, followed by the day. Press ‘Reset ’13 to view the clock at any time. 
See  page  8  if  the  time  or  date  are  incorrect.  The  clock  is  used  for  automatic 
maintenance, which takes place at 9PM every Saturday night. 

 

 

TESTING  THE  AIR  SUPPLY: 

With  the  unit  running,  press  and  hold  the  ‘Test  Air  Pressure’ 

button  14  on  the  rear  of  the  unit.  Over  a  period  of  30  seconds,  the  ‘Pressure’  10 
should  rise  to  a  minimum  of  90psi.  If  it  doesn’t  reach  90psi,  check  the  air  hose 
connections for leaks. 

 

 

DEFROST: 

Each time the unit starts up following installation, a power failure, or resetting of 

the  unit,  a  90  minute  defrost  is  commenced.  This  can  be  cancelled  at  any  time  by 
pressing  the  ‘Defrost’  button  8  on  the  front  panel.  Subsequent  defrosts  can  be 
started if required by pressing and holding the ‘Defrost’ button for 2-3 seconds.  
 

VIEWING SENSOR VALUES: 

When the system is not defrosting pressing the ‘Mode’ button 

6 will toggle the sensor value as indicated by the lights 1 

 

COOLDOWN: 

Once  defrosting  has  completed,  the  refrigeration  system  will  become  active 

and  will  begin  to  cool  down  the  liquefier.  This  can  be  monitored  by  watching  the 
“Cryo  Temp”. 

See  Viewing  Sensor  Values

.  When  the  Temp  has  reached  77K,  cool 

down is complete and production will begin. This should take about 30 minutes. 

 

LOCKING  AND  UNLOCKING: 

The  ‘Lock’  feature  9  is  a  security  device  to  stop  unwanted 

dispensing.  If  the  unit  is  unlocked  and  the  Dewar  is  in  place,  dispensing  is  enabled 
indicated by the ‘Ready’ light  2. To unlock the unit, first hold down the ‘Lock’ button 

9.  While  holding  the  ‘Lock’  button,  press  and  hold  the  ‘Mode’  button  6 

simultaneously until the lock light goes out. To relock, simply hold the ‘Lock’ button 9 
until the lock light comes on. 

Note: When locked, the unit will produce internally and 

stay  in  the  Dewar;  it  will  maintain  a  full  level  in  the  liquefier  until  the  device  is 
unlocked. Unlocking can only take place in normal operation. 
 
 
 
 

Содержание Elan2Digital

Страница 1: ...OFFICE STATION USER MANUAL MMR TECHNOLOGIES INC Volume 1 Issue 1...

Страница 2: ......

Страница 3: ...onnecting the Air Compressor Viewing Sensor Values Locking and Unlocking Front Panel Operation 8 Liquefier Level and Transfer Setting the Clock Liquid Nitrogen Removal 9 Dewar Placement Dispensing Mai...

Страница 4: ...Elan2digital systems are intended for indoor use only where clean Liquid Nitrogen is needed Because the Liquid Nitrogen generated does not get handled by complex industrial processes or trucks it is e...

Страница 5: ...Vessels MMR have thoroughly tested and approved the following Cryogenic Storage Vessels for use with this system Brymill CRY AC Series for cryosurgery MVE Dewar s from PrincetonCryo RISK OF EXPLOSION...

Страница 6: ...F Well ventilated The room should be free of dust and exhaust from other instrumentation The room should be of low humidity 70 Rel Humidity A stable dedicated power source Extension cords should be av...

Страница 7: ...ed Getting Started Getting Started CONTENTS Please ensure the following items are present Air Compressor Unit Liquefier Unit and Dewar Power Cord for Air Compressor Power Cord for Liquefier Air Compre...

Страница 8: ...r Compressor 3 Temperature LED 13 Soft Reset Unit 4 Liquefier Internal Level Indicators 14 Test Air Pressure push and hold 5 Main Display 15 Dewar Shelf 6 Toggle Sensor Values 7 Dispense Internal Cont...

Страница 9: ...s commenced This can be cancelled at any time by pressing the Defrost button 8 on the front panel Subsequent defrosts can be started if required by pressing and holding the Defrost button for 2 3 seco...

Страница 10: ...ours will be necessary before it is completely full again Flashing sequential lights indicate the unit is in Standby 11 SETTING THE CLOCK Each time the unit starts up after a power failure or the Rese...

Страница 11: ...ause the system not to recognize the Dewar and Dispensing will be disallowed DISPENSING If the unit is locked unlock it now by carrying out the procedure on page 7 Press and hold the Dispense button 7...

Страница 12: ...he unit running Step 1 Loosen the silver screw holding the filter cover in place Step 2 Partially open the cover outwards there is no need to remove the cover Step 3 Remove the dirty filter and replac...

Страница 13: ...ER 02 ER 02 ER 02 ER 02 High Humidity Error Action Check room humidity and temperature Service Air Compressor ER 03 ER 03 ER 03 ER 03 Hardware Failure Liquefier Level Sensor Error Action Contact MMR E...

Страница 14: ...nited States of America SPARE PARTS Contact our sales department at MMR for sales assistance The following parts are available for immediate purchase DESCRIPTION MMR PART 6 Foot Air Hose 1130456 006 6...

Страница 15: ...y the label at the rear of the unit Use only the cables supplied with the device and ensure a proper Ground is present Irregular unstable power may lead to hardware damage Use of equipment other than...

Страница 16: ...NS Improper handling or use of the unit along with unauthorized interference can void the warranty such as Improper handling Tipping or dropping the unit Using the unit in a way not designated in this...

Страница 17: ...r converting a gas into a liquid LN2 Abbreviation for Liquid Nitrogen LOCK The unit will produce internally until FULL but external transfers will be prohibited N2 Abbreviation for Nitrogen Gas POWER...

Страница 18: ...titive intelligence 3 3 4 Use the product in a way not described in the documentation without first obtaining written permission from MMR Engineering 4 WARRANTY 4 1 We warrant to you that the Licensed...

Страница 19: ...his End User License Agreement i Immediately at any time for any or no reason or ii If you fail to comply with any of the terms and conditions of this End User License Agreement 10 CONFIDENTIALITY 10...

Страница 20: ...MMR TECHNOLOGIES INC 1400 N Shoreline Blvd Suite A5 Mountain View California 94043 UNITED STATES OF AMERICA 1 650 962 9620 www elan2 com www mmr tech com...

Отзывы: