MMG 

TQI-021/1

 

 

 

 

Qm 

Nominal maximal flowrate for the flowmeter. 

QH 

High alarm limit for the flowrate. 

QL 

Low alarm limit for the flowrate. 

Vo 

Volume corresponding to one pulse of the remote counter. 

dT 

Pulse width of the remote counter output. 

Identifier (serial number) of the device. 

Identifier (serial number) of the flowmeter connected to the device. 

CNo 

Electronic „seal” that is the check number of the measuring system. The program of TQI-021/1 
generates this number automatically; there is no other way of setting it. In case any of the data 
in data group P is changed even temporarily, the check number will change. As there is no other 
way for changing the check number, registering and checking the value of this number can 
sense the unauthorized manipulation. 

fIn

  Selector code for input signal.  

1

 

the device receives high level pulse signal on the terminals 6(+) and 7(-), or a 6(+) and 8(-) 
(see Fig. 4) 

0

 

the device receives low level pulse signal of pick-off coil on the terminals 4 and 5, or the 
NAMUR current pulses on the terminals 8(+) and 9(-) (see Fig. 4) 

ISk

  Output current full scale. 

Attention!

 The changing of it by the user is not recommended. 

I00

  Output current zero correction. 

Attention!

 The changing of it by the user is not recommended. 

ABd 

Data of the serial line communication 

 

Form of the displayed number: 

AAA.BBBB

  

 

Where:  

 

AAA

 

digits of Address: 

BBBB

 

digits of Baud-rate: 

 

000…250 

0600

 

600 bit/s 

 

 

 

1200

 

1200 bit/s 

 

 

 

2400

 

2400 bit/s 

 

 

 

4800

 

4800 bit/s 

 

 

 

9600

 

9600 bit/s 

 

 

 

1920

 

19200 bit/s 

3.4. 

Batch control 

Setting the batch 

At pressing pushbutton 

D

 at once the value of batch (

D

)

 

appears at the display in inverse contrast 

signaling that the keys of the numeric keypad became active, enabling the entering of the new value of 
batch. After entering the new value of batch it is acknowledged by pressing pushbutton 

ENT. 

By 

pressing

 ENT

 the batching is not started, only the new value for batching is entered in memory 

D

 

Start batching by the pushbutton on front panel 

After entering the new batch value  (or leaving the old value unchanged) by pressing pushbutton 

D

 

repeatedly batching starts at once, the BATCHING information character gives a flashing 

D

 signal, the 

display is changed to displaying the batch volume left (

D

). 

 

Suspending or stop batching by front panel pushbuttons 

Batching automatically stops if the volume left (

D

)

 reaches zero. 

In case You press pushbutton 

D

 during batching started from the front panel the batching is 

suspended, the display changes to 

D

 , the value of it is seen this time in inverted contrast, signing that 

the batching is only suspended, but is not terminated. The character BATCH keeps giving the flashing 

D

 signal. In this stage batching can be continued by pressing pushbutton 

D

 again, the display of 

D

 

turns to normal contrast. The suspended batching can be terminated by pressing pushbutton 

ESC

  

Содержание TQI-021/1

Страница 1: ...U S E R M A N U A L T Q I 0 2 1 1 Flow meter signal processing unit August 2010...

Страница 2: ...ata in data group Program 6 3 4 Batch control 7 4 MECHANICAL CONSTRUCTION 8 5 SETUP 8 5 1 Unpacking the device 8 5 2 Installing the device 8 6 MAINTENANCE 8 7 REPAIR 8 8 DATA STORED IN THE MEMORY OF T...

Страница 3: ...s necessary for the calculations The TQI 021 1 signal processing unit is equipped with batch controlling function pulse output for remote counter and current output for control and recording furthermo...

Страница 4: ...able Integrated volume resettable In case of batch control the batch volume and volume left Flowrate numeric and analogue bar graph display More information of other programming data and displays see...

Страница 5: ...ctionally consists of three parts Data display Displays data set measured or calculated by the device The displayed data can consist of max 8 digits 10 5 mm high including decimal point too Informatio...

Страница 6: ...DATA GROUP is always M Only the data in group P can be rewritten The way of it is the following Switch the data group to P using the pushbutton M P Select the data to be changed by the pushbutton Pre...

Страница 7: ...AA BBBB Where AAA digits of Address BBBB digits of Baud rate 000 250 0600 600 bit s 1200 1200 bit s 2400 2400 bit s 4800 4800 bit s 9600 9600 bit s 1920 19200 bit s 3 4 Batch control Setting the batch...

Страница 8: ...want to use TQI 021 1 signal processing unit for measurement the following are recommended Study the Manual Prepare the necessary cutout on the instrument panel Figure 3 Install the device to the fina...

Страница 9: ...8 Q 2 98E 02m 3 s Actual value of flowrate M U 9 q 79 52919 Actual relative value of flowrate q Q Qm 100 M 15 I 16 728mA Actual value of output current S 16 rc0 182 5019m 3 Value of V at the moment of...

Страница 10: ...led 49 KBI can be modified S 49 KBI 0000 State of the buttons S 50 Pct 0 104875s Actual cycle time of the program S 51 Adr 1 Address of device Identification of the communication S 52 COM C BIN Commun...

Страница 11: ...level pulses during the time of dTH S 79 dTH 0s Frequency measurement time of the high level pulses S 80 DAC 2707 D A code of the current output signal S 90 D 0 Incr EXP items for factory service S 9...

Страница 12: ...Figure 2 Functions of the display Low signal limit QL High signal limit QH Flowrate Q 123456 7 Data display Information display Analog display 1 2 3 4 5 6 7 8 9 0 M P CL D ENT ESC MMG FLOW TQI 021 1 1...

Страница 13: ...t for powering transmitters Programmable current output 0 20mA max 500ohm Error contact Batch control contact Remote counter driver Low signal level flowmeter pulses 138 1 0 0 0 68 0 7 0 0 66 201 209...

Страница 14: ...4 3 2 1 10 20 11 12 13 14 15 16 17 19 18 SW1 SW2 SW3 SW4 RS232 RS485 Line powering Voltage Power Protection TQI 021 1 Sn 24VDC 10W IP20 500mA 10 9 8 7 6 5 4 3 2 1 20 11 12 13 14 15 16 17 18 19 RS232...

Страница 15: ...mplifier LA 6 1 6 Zener barrier STAHL9001 01 280 100 101 Figure 6 Connecting turbine meter sensor Figure 7 Setting of input selector on rear panel RIGHT FOR TECHNICAL CHANGES IS RESERVED 1 2 3 4 ON OF...

Страница 16: ...When 190 COM selector is in C BIN 0 or C ASC 1 state the meaning of the message bytes is the same therefore the description below refers for both cases 3 C FLOW communication protocol C BIN The devic...

Страница 17: ...de of the first character floating point 4 byte IEEE single precision PASCAL single type number first is the LOW byte of the mantissa 49H I restart of the device as it would be switched off and on Inf...

Страница 18: ...nit kg s 107 frequency unit Hz 110 no unit 111 temperature unit C 112 not used 113 density unit g l 119 voltage unit V 120 angle unit rad 121 resistance unit 150 relative unit 151 no unit 3 info byte...

Страница 19: ...he command D contains only six info bytes R or W command the value of the item VARIABLE is 1 Info byte number of the item Further info bytes depending on type bit string bits in 1 byte byte 1 byte int...

Отзывы: