MMF M68 Series Скачать руководство пользователя страница 12

10

Output

The amplifier output is buffered and DC-coupled.
Therefore, possible offset currents fed into the amplifier
output by the following equipment (for instance a PC
data acquisition board), do not cause a DC offset.

Minimum

Modulation and

Overload

Control

LEDs indicate minimum modulation and overload
condition.
The LED “>5%” lights up at an output voltage higher
than 0.7 V. The LED “OVL” lights up if the output
voltage exceeds 9 V.
The optimum gain range is selected, if the LED “>5%”
lights up and the LED “OVL” remains dark. If both
LEDs remain dark, the gain should be increased. If both
LEDs light up the gain should be reduced.
The overload detector monitors both the amplifier
output and the filter input (see Fig. 2). By that means
overload condition will also be indicated when high
signal components beyond the filter pass band occur.

)

)

)

)

An overload detector at the integrator input is not pro-
vided. In some cases high level components at higher
frequencies may overload the amplifier stage before the
integrator while at the integrator output no overload
condition can be detected. To avoid this, make sure to
check the signal level in the switch position „ACC“
(integrator off) before switching on the integrators.
When the LED indicates overload you can use the low
pass filter to attenuate high frequencies.

2.

5. Filters

Low Pass

To eliminate disturbing noise or to comply with the
Shannon theorem: “Signal frequency should be less than
half of the sampling frequency”, it can be advantageous,
to use a low pass filter. For higher accuracy in the time
domain it is recommended to set the low pass at 

1

10

the

sampling frequency.
The instruments of M68 series have 6 internal low pass
filters. The scale at the positions of the filter switch
“LOW PASS” is shows the 3 dB limiting frequencies in
kHz. The following table shows the 3 dB and the 10 %
limiting frequencies of the low pass filters:

Содержание M68 Series

Страница 1: ...Operator s Manual Metra Mess und Frequenztechnik Radebeul Meissner Str 58 D 01445 Radebeul Phone 49 351 836 2191 Fax 49 351 836 2940 Email Info MMF de Internet www MMF de Charge Amplifiers M68 Series...

Страница 2: ...duction 3 2 2 Power Supply 4 2 3 Inputs 6 2 4 Amplifier 9 2 5 Filters 10 2 6 Integrators 12 2 7 Rack Mounting Cases for Model M68R1 14 3 Technical Data 16 Appendix Limited Warranty Declaration of Conf...

Страница 3: ...ut selector switch Screws for battery compartment ICP constant current LED 5 level indicator Overload indicator Battery indicator Low pass switch High pass integrator switch Gain switch Fig 1 Front vi...

Страница 4: ...ed data acquisition systems The Signal Condi tioners provide the following functions Adaptation of the sensor signal and sensor supply Amplification High and low pass filtering for example anti aliasi...

Страница 5: ...jumper J1 in case an AC voltage shall be con nected to the input The input circuit is followed by the first amplifier stage low pass and high pass filters The low pass filter has 6 selectable limitin...

Страница 6: ...cket The positive supply terminal is connected to center pin tip The POWER ON OFF switch is located at the rear The M68R1 also has its power supply connector at the rear It is a 4 pole frame connector...

Страница 7: ...ry operation as well as for external power supply Fault Protection The instruments are protected against false polarization and short time excess voltage up to 60 V Grounding Conception The inputs and...

Страница 8: ...the so called triboelectrical effect Cables with low insulation resistance for example caused by humid connectors reduce the accuracy of measurement at lower frequencies A desirable insulation resista...

Страница 9: ...plying the sensor with constant current a positive DC voltage arises over its terminals This static bias voltage depends on manufacturer and specimen and amounts to about 5 through 14 V The sensor sig...

Страница 10: ...of Model M68R1 four screws at the side and two at the back are un screwed Jumper J1 is located left at the front side of the printed circuit board Plug it into the position OFF to switch off the cons...

Страница 11: ...this is more difficult to realize for the outputs than for the inputs because the following measuring equipment may have single ended inputs If you have the choice to use dif ferential inputs which c...

Страница 12: ...rload detector at the integrator input is not pro vided In some cases high level components at higher frequencies may overload the amplifier stage before the integrator while at the integrator output...

Страница 13: ...M68 signal conditioners have a high pass filter with a lower limiting frequency of 3 Hz 3 dB By means of this filter low frequency noise can be removed Low frequency noise may occur for example by th...

Страница 14: ...rate the measuring signal one or two times Integration can be useful for vibration measure ment by means of accelerometers on rotating machin ery Single integration of vibration acceleration results i...

Страница 15: ...n pC g and Bua is replaced by the transducer s charge sensitivity Bqa Example Vibration velocity is measured using an accelerometer with a sensitivity of Bqa 5 pC ms 2 The M68 is operated in the range...

Страница 16: ...components with high frequency and magnitude may overload the amplifier although no overload can be detected at the M68 output The over load LED remains dark This can occur due to the at tenuation of...

Страница 17: ...68B12 Channels 6 12 6 12 Built in power supply no no yes yes Fig 10 Rack case M68A6 for 6 channels The rack mounting cases without internal power sup plies Model M68A6 and M68A12 have an open rear sid...

Страница 18: ...3 Technical Data Measuring inputs Charge and ICP compatible RI 5 M BNC socket single ended ICP sensor supply 3 8 5 6 mA constant current compliance voltage 24 V switched off by internal jumper LED ind...

Страница 19: ...40 mm M68D3 7 width units x 3 height units x 190 mm M68R1 Limited Warranty Metra warrants for a period of 24 months that its products will be free from defects in material or workmanship and shall con...

Страница 20: ...d that the above mentioned products comply with the demands pursuant to the following standards EN 50081 1 EN 50082 1 EN 61000 3 EN 60950 Responsible for this declaration is the producer Metra Mess un...

Отзывы: