4.  PARAMETERS AND CALIBRATION  

4.1 Entering setting mode 

To set the parameters, turn on the weighing scales and hold the 

TARE

 key during self-

checking, until the display show “F1 CAL”. 

 

4.2 Calibration 

After entering settings mode, and setting the parameter. The display will show "F1 CAL". 
Press the 

ZERO

 

key, display shows “UNLD”, empty the platter, after the “STABLE” indicator 

turns on, press the 

ZERO

 key again. The display will show the calibrate weight. Press the 

TARE

 key to select the appropriate calibrate weight. After selection, put the calibrate weight 

on. When the 

“STABLE” indicator turns on, press the 

ZERO

 key to confirm. The display will 

shows “pass”. The calibration will be complete. 

 

4.3 Setting resolution  

After entering setting mode, the display will show 

“F1 CAL”. Press the 

TARE

 key, until the 

display shows “F2 RES”, press the 

ZERO

 key to enter, and press the 

TARE

 key to select 

the resolution. 

 

4.4 Changing capacity  

After entering the settings mode, the display will show 

“F1 CAL”. Press the 

TARE

 key until 

the 

display shows “F3 CAP”. Press the 

ZERO

 key to enter, then the 

TARE

 key to select 

capacity. Press 

ZERO

 key to confirm. 

 

4.5 Checking internal counts 

After entering the settings mode, the display will show 

“F1 CAL”. Press the 

TARE

 key until 

the 

display shows “F4 INP”. Press the 

ZERO

 key to enter, display will show internal counts. 

Press the 

TARE

 key to return settings. 

 

4.6 Automatic sleep-mode 

In weighing mode, if the scale is unmoved, and there is no change of weight, the scale will 
enter sleep mode after a minute. Increasing the weight by 4d will re-wake the scale for use. 

 

 

 

 