5-1
Series 370 Stabil-Ion Instruction Manual - 370119
Chapter 5
5
Operation of a Stabil-Ion System
The instructions in this chapter assume the instructions for Setup, Installation, and Preparing For
Operation have been completed. See Chapters 2, 3, and 4.
1.
Press the Power switch on the Control
Unit to ON (see Figure 5-1).
2.
Convectron Gauge pressures will be
displayed on lines A and B of the
display whenever the power switch is
ON. A “–” sign will be displayed if the
Convectron gauge is unplugged.
During a fast pumpdown from atmosphere, thermal effects will prevent the Convectron Gauge
from tracking pressure accurately below 10
-3
Torr. After about 15 minutes, indications in the
10
-4
range will be valid and response will be rapid. Zero adjustment at vacuum may be
performed at this time (or sooner if readings in the 10
-4
range are not needed).
Figure 5-1
Control Unit Front Panel Showing
Power ON Button
Table 5-1
Convectron Gauge Display Formats
Units
Display Format Pressure
Example
torr
scientific
< 1 Torr
3.23–3 Torr
floating point
> 1 Torr
7.15 Torr
mbar
scientific
< 1 mbar
5.18–2 mbar
floating point
> 1 mbar
88.6 mbar
scientific
> 999 mbar
1.23+3 mbar
pascal
scientific
< 1 Pa
7.23–1 Pa
floating point
> 1 Pa
78.3 Pa
scientific
> 999 Pa
1.25+4 Pa
O
I G 1
O
DEGAS
O
POWER
O
I G 2
TORR
IG
2.13 -8
A
B
5.67 -2
745
1
2
3
4
5
6
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......