3
Installation
3-19
Series 370 Stabil-Ion Instruction Manual - 370119
3.10.2 Stabil-Ion Gauge Cable Installation
Connect the Stabil-Ion cable(s) to the Power Supply first before connecting them to either the
gauges or the Control Unit.
Then connect the coaxial cable(s) to the corresponding connector on the Control Unit: IG1 to IG1,
IG2 to IG2.
Next, connect the Stabil-Ion Gauge cable(s) to the Stabil-Ion Gauge(s) as follows:
a.
Line up the locating notch on the connector locking
ring with the matching ridge on the blue connector
shell. See Figure 3-18 on page 3-19.
b.
Line up the cable connector with the gauge pin guard.
The cable will exit in line with the locating key.
c.
Insert the cable connector into the pin guard of the
Stabil-Ion Gauge.
d.
Rotate slightly until the locating key engages the cable
connector keyway.
e.
Push on the connector. Make sure the connector fully
seats and the locking ring ears engage the pin guard
ears.
f.
Rotate the blue Guard clockwise until it stops against
the detent. The cable is now locked to the Stabil-Ion
Gauge.
The Guard is designed to shield gauge surfaces that can rise to
above 100 °C. If the Guard must be removed, be sure that the
gauge is not located where personnel performing necessary
system operations might come in contact with the gauge
envelope. The connector and cable assembly must be removed
for system baking above 150 °C.
Figure 3-17
Power Supply and Control Unit Rear Views Showing the Cable Connections to the
Stabil-Ion Gauges
Figure 3-18
Stabil-Ion Gauge connector
orientation (prior to installation)
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......