2
Initial Setup Procedures
2-8
Series 370 Stabil-Ion Instruction Manual - 370119
2.1.4 RS-232 Computer Interface Setup
If your Control Unit does not have this capability, skip to Section 2.1.5 on page 2-12.
This available capability permits data output to, and gauge control by, a host computer. Output is
either by a command-response mechanism or by a talk-only mode which is invoked via a switch on
the RS-232 board. If you have this module in your unit, configure it to your system requirements by
setting the switches as directed in Section 2.1.4.1 on page 2-8.
A variety of baud rates and byte framing options are available, as well as switches to force the
handshake lines to an “always true” condition.
Stabil-Ion System RS-232 factory defaults are:
9600 BAUD, 8 data bits, no parity, 1 stop bit;
DCD, CTS, and DSR forced “true”.
The interface protocol is set using 8 switches. Refer to Figures 2-10, 2-11, and 2-12 for switch
numbering.
NOTE: Internal switches are read upon 370 power up. Changes in settings will take effect upon
next power-up cycle.
2.1.4.1 Selecting the Byte Format
2.1.4.1.1 Baud Rate
Dip switches 6-8 may be used to set the baud rate. The settings are:
Table 2-5
Baud Rate
S6
S7
S8
Baud Rate
ON
ON
ON
9600*
ON
ON
OFF
4800
ON
OFF
ON
2400
ON
OFF
OFF
1200
OFF
ON
ON
600
OFF
ON
OFF
300
OFF
OFF
ON
150
OFF
OFF
OFF
75
*
factory setting
Figure 2-10
Partial Top View of Process Control Module
X
OFF
X
ON
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......