Initial Setup Procedures
2-2
Series 370 Stabil-Ion Instruction Manual - 370119
2.1.1 Top Cover Removal
1.
With power OFF, remove any cables from Control Unit rear panel.
2.
Observe antistatic precautions to avoid damaging static sensitive components inside the
chassis. Use a grounded, conductive work surface. Do not handle integrated circuits (IC)
devices more than necessary, and only when wearing a high impedance ground strap.
(A high impedance helps protect human life in case of inadvertent contact with high voltage.)
3.
Remove the four Phillips head screws identified in Figure 2-2. If the unit is equipped with a
rear bracket instead of one rear screw, unfasten the Phillips head screw on the bracket, and
slide off the bracket
4.
Lift off top cover.
Figure 2-2
Location of Screws for Removal of Top Cover