7
Service and Maintenance
7-12
Series 370 Stabil-Ion Instruction Manual - 370119
6.
Carefully remove the bus ribbon cable from all modules located to the right (as you face the
front panel) of the position where the module is to be installed. Remove connectors slowly
using pull tabs.
7.
Lift out the filler module at the position where the module is to be installed.
8.
Install the module in its proper position making sure all ends lock together.
9.
Carefully reconnect the bus ribbon connectors.
10.
Select appropriate switch settings. See Chapter 2.
11.
Replace the top cover as directed in Section 2.1.6 on page 2-14.
7.15 Troubleshooting Flow Charts
Table 7-10
Troubleshooting Flow Charts
Fig. No.
Description
Fig. No.
Description
Figure 7-6
Power Turn ON
NO
YES
YES
370 TURN ON
SEQUENCE BEGIN
Set switches as follows:
POWER SWITCH
ON
IS
POWER SWITCH
LAMP ON?
IS
ION GAUGE
PRESSURE
DISPLAY
BLANK?
ARE
IG GAUGE
#1 AND #2
ON/OFF
LAMPS
OFF?
ARE
CONVECTRON
A & B
DISPLAYS
ON?
370 TURN ON
SEQUENCE FINISH
TURN OFF
POWER SWITCH
CHECK AC
LINE FUSE
REPLACE FUSE
IS
FUSE
BLOWN?
CHECK AC
LINE CORD
CHECK MAIN
CIRCUIT BREAKER
TOGGLE
CIRCUIT BREAKER
REPLACE CORD
IS
AC LINE CORD
DAMAGED?
IS
CIRCUIT
BREAKER
TRIPPED?
IS
CONTR.
INTERCONNECT
CABLE
DAMAGED?
ARE
SWITCHES
NOT SET
CORRECTLY?
CHECK POWER
SUPPLY AC
LINE VOLTAGE
SWITCHES
CHECK
CONTROLLER
INTERCONNECT
CABLE
CALL FACTORY
REPLACE CABLE
SET SWITCHES
TO CORRECT
VOLTAGE
NO
NO
NO
NO
NO
NO
NO
YES
YES
YES
YES
YES
YES
YES
NO
IG1 and IG2 auto: off
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......