Operation
Mini-Convectron Module with RS-485
31
Instruction Manual - 275545
Saf
e
ty Ins
tructions
Introduction
Installatio
n
O
p
e
ra
tion
M
a
inten
ance
Calibration
TS
Set span (typically at atmospheric pressure)
Use scientific notation.
Example:
From host:
#01TS7.60E+02 CR
From Mini-Convectron:
*01_PROGM_OK CR
Possible Responses:
*01_HI_ATM_V CR
Transducer output voltage is higher
than it should be. Actual pressure
may be higher or there may be tube
contamination which may cause
readout error to increase between
TZ and TS settings. If
?01_OFST_LIM CR was received
while setting TZ, this response may
occur.
*01_LO_ATM_V CR
Transducer output voltage is lower
than it should be. Similar to
HI_ATM_V.
?01_GAIN_LIM CR
Gain programmed at limit. Readout
errors will occur even at TS setting.
?01_DEF_SNSR CR
Sensor defect, no change in
programming, see maintenance
section.
?01_RANGE_ER CR
Command error, TS must be set
above 399 Torr. No change in
programming.
Notes:
Do this only at higher pressures. If performed at vacuum, an error message
response will occur. The change occurs as soon as the function is
performed.