25 | ESP302-GPIB-ADPT
A1387A0-EDH0429En1010 11/21
USER's MANUAL
6 Appendix
6.1 Self-Test Error Codes
At power turn on, the ESP302-GPIB-ADPT conducts a self-test of its major components. A successful
test ends when the RDY LED on. Test failures are indicated by the LED patterns shown in Table
below. Typically, the failure is a pattern of blinking LEDs which lasts for 5 seconds. If the ESP302-
GPIB-ADPT does not have a fatal error, the self-test will continue until it is complete but the RDY LED
will remain off. If a self-test failure occurs, turn the unit off for 10 seconds and turn power back on. If
the failure persists, refer to paragraph 5.3 for repair instructions.
Table below lists the ESP302-GPIB-ADPT self-test error codes and most probable faulty component.
Table 3: ESP302-GPIB-ADPT Self-Test Error Codes
Front Panel LED
Fault
Resp.
PWR
RDY
TLK
LSTN
SQR
ERR
-
⊕
-
-
-
-
-
fatal error (CPU, Flash RAM …etc.)
-
-
-
-
-
-
-
fatal error (power supply)
-
⊕
⊕
⊕
⊕
⊕
⊕
fatal error (CPU, Flash RAM …etc.)
4
⊕
-
-
B
-
-
7210-U13
0
⊕
⊕
x
x
-
-
Self-test passed successfully
LED legend:
⊕
led is on
B led is blinking
- led is off
x led could be on or off