6
Service & Maintenance
33
Series 907 Analog Convection Transducer
Instruction Manual - #109070001
5.
Adjust potentiometer R38 to a voltmeter reading of 5.487 V.
6.
Calibrate the sensor for vacuum and atmosphere following the steps on page 14.
7.
Disconnect the power cable.
6.10.2 Reassemble the Electronics Subassembly
1.
Carefully replace the front cover of the case.
2.
Secure the remaining sensor tube screw and the screw lock.
6.11 Bakeout the Sensor Tube
See Section 6.10 to remove the transducer electronics case.
1.
Remove the two screws holding the sensor tube to the transducer electronics subassembly
with a #1 Phillips head screwdriver. Separate the electronics from the sensor.
2.
Bake the standard sensor tube to a maximum of 200°C and the CE marked model to a
maximum of 100°C.
3.
Reassemble the transducer electronics on the sensor (see Section 6.10.2).
Hot Surface
Do not touch the gauge or the immediate surrounding area of
the vacuum chamber during the bake out process. Wait until
the bake out process has finished and the area has cooled.