40
A
Electromagnetic Compatibility
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND OPERATING TABLE MOT-1602BW
Operating Table MOT-1602BW is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of Operating Table MOT-1602BW can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and Operating Table MOT-1602BW as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
W
d = 1.2 P
d = 1.2 P
d = 2.3 P
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where
P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1
: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by