No. 99MAM016B
iv
Precautions for Use
IMPORTANT
To ensure safe operates with this instrument, check the following items prior to
use.
The Input Tool operates with the power supplied form the PC connected. De-
pending on the manufacturers or models, PCs may generate significant inter-
ference noise. Be sure to check that the entire system operates normally before
starting measurement. Check the system according to the following procedure.
1)Connect the Input tool to measuring tool and the PC (or equivalent),then
turn the PC on.(It needs not to open the data input window.)
2)Zero set the measuring tool.(Refer to the respective User's Manual)
3)Open the measuring faces(contact point)to the point of workpiece insertion,
then close them to the original zero point. Check to be sure that the meas-
uring tool displays zero.
4)Repeat step 3 for 5 times. If the measuring tool repeats zero for each time,
you may safely use the system for measurement.
5)In the event that the instrument malfunction should result, please contact
Mitutoyo office.
CAUTION
・
If a malfunction is not solved with switches, remove the cables and connect
them again.
・
Wait more than one second before outputting the data with the “Input Tool for
RS232C” after the completion of previous data output to avoid outputting an
error data.
Fig.1 Do not use this product at site where it may be subjected to water or oil.
Fig.2 Do not leave this product at site where it may be subjected to direct
sunlight, in the car, near the heat or extremely cold places.
Fig.3 Do not drop this product.
Fig.1 Fig.2 Fig.3