APPENDICES
Appendix 1 Accuracy
41
A
*1 If a value outside the temperature measuring range in the table is input from a thermocouple, the value is regarded as the maximum
value or minimum value of the temperature measuring range.
*2 The accuracy is applied to only the temperature range of class 1 to 3 of JIS C 1602-1995.
*3 The accuracy is not guaranteed though temperature measurement is possible.
*4 The value indicates a temperature error generated per wiring resistance 1
of a thermocouple. Check the resistance value and
calculate the temperature error of the system. If the error is beyond the allowable range of the system used, correct the measured
temperature value with the offset/gain setting.
J
-210 to -40
-40 to 750
0.2
0.06
, or a larger value of -0.02%
or +0.02% of measured temperature
3.95
0.006
/
750 to 1200
T
-270 to -200
-200 to 0
0.5
0.06
, or a larger value of -0.1% or
+0.1% of measured temperature
5.5
0.016
/
0 to 350
0.4
0.06
, or a larger value of -0.02%
or +0.02% of measured temperature
2.15
0.007
/
350 to 400
N
-270 to -200
-200 to 0
0.5
0.06
, or a larger value of -0.2% or
+0.2% of measured temperature
6.2
0.025
/
0 to 1250
0.5
0.06
, or a larger value of -0.02%
or +0.02% of measured temperature
6.75
0.01
/
1250 to 1300
Usable
thermocouple
Temperature
measuring
range
*1
Conversion accuracy
(for operating ambient
temperature 25
5
)
Temperature characteristics
(for 1
of operating ambient
temperature change)
Maximum
temperature
error at ambient
temperature 55
Effect per wiring
resistance 1
*4