background image

13

mvip@mitech-ndt.com

3. Use the

key and the

key to select the desired file.

4. Press the

key to print the selected file. This operation will send all the data in current file to the

mini printer via RS232 port and print them out.

5. Press the

key to exit the data logging functions and return to measurement mode.

4.10 Beep Mode

When the beep is set to

On

】,

it would make a short hoot while press the key each time, on each

measurement, or the measured value exceeds the tolerance limit.

Press the

key to switch the beep mode on and off. It will display the current beep mode on the

main screen.

4.11 EL Backlight

With the background light, it is convenient to work in the dark condition. Press key

to switch

on or switch off the background light at any moment as you need after power on. Since the EL light will
consume much power, turn on it only when necessary.

4.12 Battery Information

Two AA size alkaline batteries are needed as the power source. After several hours’ usage of the

preset batteries, the battery symbol on the screen will be shown as

. The more of dark part

indicates the more close to fill. When the battery capacity runs out, the battery symbol will be shown
as

and will begin to flash. When this occurs, the batteries should be replaced.

Please take out the batteries when not working during a long period of time.

4.13 Auto Power Off

The instrument features an auto power off function designed to conserve battery life. If the tool is

idle for 5 minutes, it will turn itself off. While the voltage of the battery is too low this function will also
work.

4.14 System Reset

Press down the

key while powering on the instrument will restore factory defaults. All the

memory data will be cleared during system reset. The only time this might possibly helpful is if the
parameter in the gauge was somehow corrupted.

4.15 Connecting to a Computer

MT160 is equipped with a RS232 serial port. Using the accessory cable, the gauge has the ability

to connect to a computer, or external storage device. Measurement data stored in the memory of the
gauge can be transferred to the computer through the RS232 port. Detailed information of the
communication software and its usage refer to the software manual.

5 Servicing

When the hardness tester appears some other abnormal phenomena, please do not dismantle or

adjust any fixedly assembled parts. Fill in and present the warranty card to us. The warranty service
can be carried on.

6 Transport and Storage

Keep it away from vibration, strong magnetic field, corrosive medium, dumpiness and dust.
Storage in ordinary temperature.

With original packing, transport is allowed on the third grade highway.

[email protected]

ponpe.com

Содержание MT150

Страница 1: ...Ultrasonic Thickness Gauge MT150 160 User s Manual MITECH CO LTD www mitech ndt com www ponpe com ...

Страница 2: ...urfaces 8 4 Operation 8 4 1 Power On Off 8 4 2 Probe Zero 8 4 3 Sound Velocity Calibration 9 4 4 Making Measurements 10 4 5 Scan mode 11 4 6 Changing Resolution 11 4 7 Changing Units 11 4 8 Memory Management 12 4 9 Data Printing 12 4 10 Beep Mode 13 4 11 EL Backlight 13 4 12 Battery Information 13 4 13 Auto Power Off 13 4 14 System Reset 13 4 15 Connecting to a Computer 13 5 Servicing 13 6 Transpo...

Страница 3: ... 300 mm in Steel Sound Velocity Range 1000 9999 m s Resolution MT150 0 1mm MT160 0 1mm 0 01mm Accuracy 0 5 Thickness 0 04 mm depends on materials and conditions Units Metric Imperial unit selectable Four measurements readings per second for single point measurement and ten per second for Scan Mode Memory for up to 20 files up to 99 values for each file of stored values Power Source Two AA size 1 5...

Страница 4: ...al ultrasonic thickness gauge determines the thickness of a part or structure by accurately measuring the time required for a short ultrasonic pulse generated by a transducer to travel through the thickness of the material reflect from the back or inside surface and be returned to the transducer The measured two way transit time is divided by two to account for the down and back travel path and th...

Страница 5: ...een Coupling Status Indicate the coupling status While the gauge is taking a measurement the coupling status should be on If it is not on or not stable the gauge is having difficulty achieving a stable measurement and the thickness value displayed will most likely be erroneous Unit Current unit system MM or IN for thickness value M S or IN μS for sound velocity Battery Information Display the rest...

Страница 6: ... being measured than a small emitting area Thus a so called 1 2 inch transducer will emit a stronger signal than a 1 4 inch transducer Absorption and Scattering As ultrasound travels through any material it is partly absorbed If the material through which the sound travels has any grain structure the sound waves will experience scattering Both of these effects reduce the strength of the waves and ...

Страница 7: ...transducer The two semicircles of the wearface are visible as is the barrier separating them One of the semicircles is responsible for conducting ultrasonic sound into the material being measured and the other semicircle is responsible for conducting the echoed sound back into the transducer When the transducer is placed against the material being measured it is the area directly beneath the cente...

Страница 8: ...orn on one side more than another the sound beam penetrating the test material may no longer be perpendicular to the material surface In this case it will be difficult to exactly locate tiny irregularities in the material being measured as the focus of the sound beam no longer lies directly beneath the transducer 4 Operation 4 1 Power On Off The instrument is turned on by pressing the key The gaug...

Страница 9: ...r to calibration will result in a multi material velocity calculation that may be different from the actual material velocity intended to be measured 4 3 1 Calibration to a known thickness Note This procedure requires a sample piece of the specific material to be measured the exact thickness of which is known e g from having been measured by some other means 1 Perform a Probe Zero 2 Apply couplant...

Страница 10: ... changes from the value shown while the transducer was coupled repeat step 3 5 Press the key The MM or IN symbol should begin flashing 6 Use the key and the key to adjust the displayed thickness up or down until it matches the thickness of the sample piece 7 Press the key The display will flash 1OF2 Repeat steps 3 through 6 on the second calibration point 8 Press the key so that The M S or IN μS s...

Страница 11: ... solely with the user of the instrument 4 5 Scan mode While the gauge excels at making single point measurements it is sometimes desirable to examine a larger region searching for the thinnest point The gauge includes a feature called Scan Mode which allows it to do just that In normal operation the gauge performs and displays four measurements every second which is quite adequate for single measu...

Страница 12: ...ements 3 Press the key on the desired file It will automatically clear the file and display DEL 4 Press the key at any time to exit the data logging functions and return to measurement mode 4 8 3 Viewing deleting stored record This function provides the user with the ability to view delete a record in a desired file previously saved in memory Following is the steps 1 Press the key to activate the ...

Страница 13: ...hould be replaced Please take out the batteries when not working during a long period of time 4 13 Auto Power Off The instrument features an auto power off function designed to conserve battery life If the tool is idle for 5 minutes it will turn itself off While the voltage of the battery is too low this function will also work 4 14 System Reset Press down the key while powering on the instrument ...

Страница 14: ...ss 0 222 5639 Rubber vulcanized 0 091 2311 Teflon 0 056 1422 Water 0 058 1473 Appendix B Applications Notes Measuring pipe and tubing When measuring a piece of pipe to determine the thickness of the pipe wall orientation of the transducers is important If the diameter of the pipe is larger than approximately 4 inches measurements should be made with the transducer oriented so that the gap in the w...

Страница 15: ...efore sound velocity may vary considerably from one piece to another Some laminated materials may even exhibit noticeable changes in sound velocity across a single surface The only way to reliably measure such materials is by performing a calibration procedure on a sample piece of known thickness Ideally this sample material should be a part of the same piece being measured or at least from the sa...

Страница 16: ...tures may include water various oils and greases gels and silicone fluids Measurements at elevated temperatures will require specially formulated high temperature couplants Inherent in ultrasonic thickness measurement is the possibility that the instrument will use the second rather than the first echo from the back surface of the material being measured while in standard pulse echo mode This may ...

Страница 17: ...ing to repair sheet In accordance with the international relevant regulations the following are not within the scope of free warranty Damage caused by man made or improper keeping Self dismantle or non special repair shop dismantle Do not follow the requirement of service registration or warranty expired Consumable parts Service promise MITECH users have lifelong maintenance service Free maintenan...

Отзывы: