Warning: When the battery pack is being charged, printing or paper feeding is unavailable.
6.16 Battery Replacement
When the battery pack fails to be charged, the user should replace the batteries following the program
below:
Power down the instrument.
Take off the battery compartment cover and take out the battery pack.
Insert the connection plug of the new battery pack into the socket on the circuit board.
Reset the battery cover.
Turn on the instrument to check.
Warning: Please pay much attention to the polarity of the battery during battery replacement.
6.17 Connection of Data Transmission Cable
Insert one connection plug of the transmission cable into the USB socket on the right side of main body,
and insert the another plug into the USB port on the back of computer box. Refer to the manual of the
DataPro software for detailed information.
7 Fault Analysis & Evacuation
Fault Appearance
Fault Analysis
Handling method
Charge failure
Battery failure
Replace the battery with a
new pack
The battery is switched off
Switch on the battery
No measured value
Impact device cable failure
Replace the cable
Failure power on
Battery exhaustion
Charge the battery
The battery is switched off
Switch on the battery
8 Servicing & Maintenance
8.1 Impact Device Servicing
After the impact device has been used for 1000--2000 times, please use the nylon brush provided to
clean the guide tube and impact body. When cleaning the guide tube, unscrew the support ring first, then
take out the impact body, spiral the nylon brush in counter-clock direction into the bottom of guide tube
and take it out for 5 times, and then install the impact body and support ring again.
Release the impact body after use.
Any lubricant is absolutely prohibited inside the impact device.
8.2 Normal Maintenance Program
When using standard Rockwell hardness block to testing, if all the error is bigger than 2 HRC, it may be
the invalidation of impacted ball top caused by abrasion. Changing the spherical test tip or impact object
should be considered.
When the hardness tester appears some other abnormal phenomena, please do not dismantle or adjust
any fixedly assembled parts. Fill in and present the warranty card to us. The warranty service can be
carried on.
9 Calibration
Calibration is needed every 1 year.
10 Notice of Transport and Storage Conditions
Keep it away from vibration, strong magnetic field, corrosive medium, dumpiness and dust. Storage
in ordinary temperature.
With original packing, transport is allowed on the third grade highway.