9.1 Physical Characteristics
Compliant with IEC 61526 Ed. 3, ANSI 42.20 (*)
(
*) Isotropy at
241
Am and
137
Cs with angle ± 75°
Measurement and display
:
X-ray and gamma energy range: 15 keV to 7 MeV
Energy response better than ± 20% from 16 keV to 7 MeV
(typically ± 10%)
Accuracy
H
p(10)
:
≤
± 10% (16 keV)
≤
± 10% (
241
Am)
≤
± 5% (
137
Cs)
≤
± 10% (
60
Co)
Display units
: mSv, µSv, or mrem
Display dose
: 1 µSv to 10 Sv (0.1 mrem to 1000 rem)
Display rate
:
0.01 mSv/h to 10 Sv/h (1 mrem/h to 1000 rem/h)
or 0.001 mSv/h to 10 Sv/h (0.1 mrem/h to 1000 rem/h)
(extended option(**))
Measurement range
:
1 µSv to 10 Sv (0.1 mrem to 1000 rem)
0.1 µSv/h to 20 Sv/h (0.01 mrem/h to 2000 rem/h)
Saturation indication : above 10 Sv (1000 rem)
or 10 Sv/h (1000 rem/h)
Dose rate linearity
:
≤
± 20 % up to 10 Sv/h (1000 rem/h) (Co and X H30 20 keV)
(**) - Mandatory if rate alarms are set below 50 µSv/h
(5 mrem/h) in order to display rate with better accuracy, if
fast response time for rate alarm is not required.
- Recommended for rate alarms settings below
100 µSv/h (10 mrem/h), if fast response time for rate alarm
is not required.
Содержание DMC 3000
Страница 1: ...151153EN G DMC 3000 User s Guide Health Physics Division Radiation Safety Perfected 151153EN G...
Страница 16: ...16 151153EN G DMC 3000 User s Guide States Modes Long press 2s T O Time Out 4 1 Run Pause Flow Diagram...
Страница 46: ...46 151153EN G DMC 3000 User s Guide 12 CE Certificate...
Страница 48: ...48 151153EN G DMC 3000 User s Guide Notes...
Страница 49: ...49 151153EN G DMC 3000 User s Guide Notes...
Страница 50: ...50 151153EN G DMC 3000 User s Guide Notes...
Страница 51: ...51 151153EN G DMC 3000 User s Guide...