Appendix C Power-On Error Codes
C. Power-On Error Codes
When the 7401 is turned on or reset from the front panel (
Initializing the Display
on
page 9), the microprocessor runs several tests. If an error is detected one of the
following error codes will be displayed:
E1
RAM test error. RAM is on board the micro-processor.
E2
Checksum error for program code in EPROM on board the microprocessor.
E3
A/D error. No change in A/D status line within a 100 ms time period.
E4
The ac power line was not detected by the module. Module will use internal
timer as time base.
E5
The ac power line was detected by the module but the line frequency is not
within ± 0.5 Hz of 50 or 60 Hz. Module will use internal timer as time base.
26
7401/7401VR Alpha Spectrometer User’s Manual - 9231603C