Regulatory Information
Regulatory Information
55
54
The Miracle-EarCHARGE
™
start RIC/BTE AV charger has been tested to, and has
passed, the following emissions and immunity tests:
• IEC 60601-1-2 radiated and conducted emissions requirements for a Group 1
Class B device as stated in CISPR 11.
• Harmonic distortion and voltage fluctuations affecting the power input source as
stated in Table 2 of IEC 60601-1-2.
• RF radiated immunity at a field level of 10 V/m between 80 MHz and 2.7 GHz as
well as higher field levels from communications devices as stated in Table 9 of IEC
60601-1-2.
• Immunity to power frequency magnetic fields at a field level of 30 A/m and
proximity magnetic fields as stated in Table 11 of IEC 60601-1-2.
• Immunity to ESD levels of +/- 8 kV conducted discharge and +/- 15 kV air
discharge.
• Immunity to Electrical fast transients on the power input at a level of +/- 2 kV at a
100 Hz repetition rate.
• Immunity to Surges on the power input of +/- 1 kV line to line.
• Immunity to conducted disturbances induced by RF fields on the power input as
stated in Table 6 of IEC 60601-1-2.
• Immunity to voltage dips and interruptions on the power input as stated in Table 6
of IEC 60601-1-2.
WIRELESS NOTICES
FCC ID: EOA-24LIVIORCHG
IC: 6903A-24LIVIORCHG
FCC Notice
This hearing aid and Miracle-EarCHARGE
™
start RIC/BTE AV charger comply with part
15 of the FCC rules and with ISED Canada license-exempt RSS standards. Operation
is subject to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation of the device.
NOTE:
The manufacturer is not responsible for any radio or TV interference caused
by unauthorized modifications to this equipment. Such modifications could void the
user’s authority to operate the equipment.